A-series lift - A625 Manual Part No. 324117. Prism Medical.
Published March 15, 2017 · Recall initiated: January 30, 2017
Summary
FDA records medical device Class II recall Z-1406-2017, published March 15, 2017 and initiated January 30, 2017. Recalling firm: ErgoSafe Products, LLC (DBA) Prism Medical; Status: Terminated; Reported quantity: 76; Lot: A6MC000016 A6MC000019 A6MC000020 A6MC000021 A6MC000022 A6MC000023 A6MC000024 A6MC000025 A6MC000026 A6MC000027 A6MC000028 A6MC000029 A6MC000030 A6MC000031 A6MC000009 A6MC000002 A6MC000004 A6MC000001 A6MC000003 A6MC000005 A6MC000006 A6MC000007 A6MC000008 A6MC000010 A6MC000011 A6MC000012 A6MC000013 A6MC000014 A6MC000015 A6MC000017 A6MC000018 A6MC000096 A6MC000087 A6MC000088 A6MC000090 A6MC000091 A6MC000092 A6MC000093 A6MC000094 A6MC000048 A6MC000068 A6MC000070 A6MC000072 A6MC000073 A6MC000074 A6MC000075 A6MC000078 A6MC000079 A6MC000081 A6MC000082 A6MC000083 A6MC000085 A6MC000086 A6MC000042 A6MC000043 A6MC000045 A6MC000046 A6MC000047 A6MC000049 A6MC000050 A6MC000051 A6MC000052 A6MC000053 A6MC000054 A6MC000055 A6MC000056 A6MC000057 A6MC000058 A6MC000059 A6MC000060 A6MC000061 A6MC000062 A6MC000063 A6MC000064 A6MC000066 A6MC000071. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- ErgoSafe Products, LLC (DBA) Prism Medical
- Distributor
- Product was distributed to MT and NV; one VA consignee.
- Lot
- A6MC000016 A6MC000019 A6MC000020 A6MC000021 A6MC000022 A6MC000023 A6MC000024 A6MC000025 A6MC000026 A6MC000027 A6MC000028 A6MC000029 A6MC000030 A6MC000031 A6MC000009 A6MC000002 A6MC000004 A6MC000001 A6MC000003 A6MC000005 A6MC000006 A6MC000007 A6MC000008 A6MC000010 A6MC000011 A6MC000012 A6MC000013 A6MC000014 A6MC000015 A6MC000017 A6MC000018 A6MC000096 A6MC000087 A6MC000088 A6MC000090 A6MC000091 A6MC000092 A6MC000093 A6MC000094 A6MC000048 A6MC000068 A6MC000070 A6MC000072 A6MC000073 A6MC000074 A6MC000075 A6MC000078 A6MC000079 A6MC000081 A6MC000082 A6MC000083 A6MC000085 A6MC000086 A6MC000042 A6MC000043 A6MC000045 A6MC000046 A6MC000047 A6MC000049 A6MC000050 A6MC000051 A6MC000052 A6MC000053 A6MC000054 A6MC000055 A6MC000056 A6MC000057 A6MC000058 A6MC000059 A6MC000060 A6MC000061 A6MC000062 A6MC000063 A6MC000064 A6MC000066 A6MC000071
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAColor contact lenses labeled under the following brands: Diamond, Fierce, Ambition, Circle Collection,March 15, 2017
- FDAMicroAire K-WiresMarch 15, 2017
- FDAAngioJet® Ultra 5000 A Console:Foot Switch Material number (U PN) 105650 Cardiovascular: The Console is aMarch 15, 2017
- FDACarry bar which is an accessory to a lift used for patients in hospitals and other facilities. Recall impactsDecember 21, 2016
- FDAPrism Medical P-300 Portable Consumer Lift; PN 303050, 303050R, and 303051; weight specification 300 lbs. /136November 25, 2015
- FDAPrism Medical C-300 Consumer Ceiling Lift; PN 330050, 330050R, 330053, 330053R, and 330057; weightNovember 25, 2015
Questions about this recall
Was A-series lift - A625 Manual Part No. 324117. Prism Medical. recalled?
Yes. FDA published recall Z-1406-2017 on March 15, 2017. Reported hazard: During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.
What should I do if I have this product?
Voluntary: Firm initiated · Letter