Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and stabilization of spinal segments.
Published June 5, 2013 · Recall initiated: May 6, 2013
Summary
FDA records medical device Class II recall Z-1421-2013, published June 5, 2013 and initiated May 6, 2013. Recalling firm: Alphatec Spine, Inc.; Status: Terminated; Reported quantity: 1,814; Lot: 83010 85001 85002 83005-30 83005-35 83005-40 83005-45 83005-50 83006-30 83006-35 83006-40 83006-45 83006-50 83006-55 83006-60 83007-30 83007-35 83007-40 83007-45 83007-50 83007-55 83008-30 83008-35 83008-40 83008-45 83008-50 83008-55 83008-80 82001-1 Catalog Numbers: LA-HSS0010 LA-HDOO01 LA-CTOO02 LA-PS0530 LA-PS0535 LA-PS0540 LA-PS0545 LA-PS0550 LA-PS0630 LA-PS0635 LA-PS0640 LA-PS0645 LA-PS0650 LA-PS0655 LA-PS0660 LA-PS0730 LA-PS0735 LA-PS0740 LA-PS0745 LA-PS0750 LA-PS0755 LA-PS0830 LA-PS0835 LA-PS0840 LA-PS0845 LA-PS0850 LA-PS0855 LA-PS0880 LA-SOO01 Lots Affected: All. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head and rod may fail to provide a rigid construct under loading. To date there have been no reports of patient injury, adve
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alphatec Spine, Inc.
- Distributor
- Nationwide Distribution.
- Lot
- 83010 85001 85002 83005-30 83005-35 83005-40 83005-45 83005-50 83006-30 83006-35 83006-40 83006-45 83006-50 83006-55 83006-60 83007-30 83007-35 83007-40 83007-45 83007-50 83007-55 83008-30 83008-35 83008-40 83008-45 83008-50 83008-55 83008-80 82001-1 Catalog Numbers: LA-HSS0010 LA-HDOO01 LA-CTOO02 LA-PS0530 LA-PS0535 LA-PS0540 LA-PS0545 LA-PS0550 LA-PS0630 LA-PS0635 LA-PS0640 LA-PS0645 LA-PS0650 LA-PS0655 LA-PS0660 LA-PS0730 LA-PS0735 LA-PS0740 LA-PS0745 LA-PS0750 LA-PS0755 LA-PS0830 LA-PS0835 LA-PS0840 LA-PS0845 LA-PS0850 LA-PS0855 LA-PS0880 LA-SOO01 Lots Affected: All
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix SystemJuly 29, 2026
- FDAIdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA GraftMarch 6, 2024
- FDAAtec Insignia Anterior Cervical Plate System, REF 136-0354 Insignia, ACP, 3-Level, 54 mm, Rx Only,June 16, 2021
- FDAAtec Insignia Anterior Cervical Plate System, REF 136-0248 Insignia, ACP, 2-Level, 48 mm, Rx Only,June 16, 2021
- FDARayStation Therapy Treatment Planning System, Model Numbers 2.0, 2.5, and 3.0; CLASSIFICATION NAME: System,June 5, 2013
- FDABBL DrySlide PYR Kit, catalog number 231747 Product Usage: Presumptive diagnostic aid for group A streptococciJune 5, 2013
Questions about this recall
Was Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and stabilization of spinal segments. recalled?
Yes. FDA published recall Z-1421-2013 on June 5, 2013. Reported hazard: Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head and rod may fail to provide a rigid construct under loading. To date there have been no reports of patient injury, adve
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit