PRIMUS HI and Mevatron M2/Primus Mid-Energy, Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer
Published March 15, 2017 · Recall initiated: January 25, 2017
Summary
FDA records medical device Class II recall Z-1432-2017, published March 15, 2017 and initiated January 25, 2017. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 35 systems; Lot: Device Model Number 4504200 and 1940035 Serial numbers: 3009 2958 2863 2972 2646 3079 2516 2962 3055 1452 2702 2125 3066 2689 2228 2564 2507 1904 3075 2375 3013 3068 2596 3025 3049 2965 2045 3084 2855 2906 2292 2384 2067 2818 3089. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Nationwide Distribution
- Lot
- Device Model Number 4504200 and 1940035 Serial numbers: 3009 2958 2863 2972 2646 3079 2516 2962 3055 1452 2702 2125 3066 2689 2228 2564 2507 1904 3075 2375 3013 3068 2596 3025 3049 2965 2045 3084 2855 2906 2292 2384 2067 2818 3089
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAMerge PACS software. The firm name on the label is Merge Healthcare.March 15, 2017
- FDAMEVATRON M 6730/6740 (MEX) Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS familyMarch 15, 2017
Questions about this recall
Was PRIMUS HI and Mevatron M2/Primus Mid-Energy, Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to deliver x-ray radiation for therapeutic treatment of cancer recalled?
Yes. FDA published recall Z-1432-2017 on March 15, 2017. Reported hazard: Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
What should I do if I have this product?
Voluntary: Firm initiated · Letter