Centurion hook kits labeled as: a) STERILE SKIN HOOK FINE (EA3975F), Product Code 66660; b) STERILE SMALL 2 PRONG SKIN HOOK (25754), Product Code 67105; c) STERILE NEWS TRACH HOOK(83300), Product Code 67810; d) STERILE LILLIE EAR HOOK DELICATE 7", Product Code I68655
Published April 10, 2024 · Recall initiated: February 23, 2024
Summary
FDA records medical device Class II recall Z-1435-2024, published April 10, 2024 and initiated February 23, 2024. Recalling firm: MEDLINE INDUSTRIES, LP - Northfield; Status: Ongoing; Reported quantity: 1520 units; Lot: a) 66660, UDI/DI 60653160052154 (case) 00653160052152 (each), Lot Numbers: 2023042790, 2023091190; b) 67105, UDI/DI 60653160061880 (case) 00653160061888 (each), Lot Numbers: 2023042690; c) 67810, UDI/DI 30653160196703 (case) 00653160196702 (each), Lot Numbers: 2023060790, 2023070790; d) I68655, UDI/DI 00653160317220 (case) 10653160317227 (each), Lot Numbers: 2023061290. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Distributor
- Worldwide distribution - US Nationwide and the countries of Panama, Canada.
- Lot
- a) 66660, UDI/DI 60653160052154 (case) 00653160052152 (each), Lot Numbers: 2023042790, 2023091190; b) 67105, UDI/DI 60653160061880 (case) 00653160061888 (each), Lot Numbers: 2023042690; c) 67810, UDI/DI 30653160196703 (case) 00653160196702 (each), Lot Numbers: 2023060790, 2023070790; d) I68655, UDI/DI 00653160317220 (case) 10653160317227 (each), Lot Numbers: 2023061290
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACURAD Alcohol Prep Pads, Sterile, Medium, 2-Ply, Contents: 5 boxes per Carton, 30 Boxes per Case, Single UseSeptember 3, 2025
- FDAReliOn, Sterile Alcohol Swabs, Skin Cleanser, 200 Swabs, Distributed by, Walmart, Inc., Bentonville, AR 72716September 3, 2025
- FDAGood Neighbor Pharmacy, Alcohol Prep Pads, Distributed By AmeriSource Bergen, 1 West First Avenue,September 3, 2025
- FDAMedline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004ASeptember 3, 2025
- FDACenturion retractor kits labeled as: a) STERILE 6-1/2" SENN RETRACTOR (ST7185), Product Code 66795; b) STERILEApril 10, 2024
- FDACenturion scalpel handle kits labeled as: a) STERILE # 3 SCALPEL HANDLE, Product Code 66825; b) STERILE PAS #3April 10, 2024
Questions about this recall
Was Centurion hook kits labeled as: a) STERILE SKIN HOOK FINE (EA3975F), Product Code 66660; b) STERILE SMALL 2 PRONG SKIN HOOK (25754), Product Code 67105; c) STERILE NEWS TRACH HOOK(83300), Product Code 67810; d) STERILE LILLIE EAR HOOK DELICATE 7", Product Code I68655 recalled?
Yes. FDA published recall Z-1435-2024 on April 10, 2024. Reported hazard: Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit