Centurion manual surgical kits labeled as: a) STERILE ALLIS TISS FCP 10", Product Code I68905; b) VAGINAL REPAIR KIT, Product Code MNS11505; c) PTCA ADD-ON KIT, Product Code MNS11545; d) INSTRUMENT SET, Product Code SUT21355
Published April 10, 2024 · Recall initiated: February 23, 2024
Summary
FDA records medical device Class II recall Z-1443-2024, published April 10, 2024 and initiated February 23, 2024. Recalling firm: MEDLINE INDUSTRIES, LP - Northfield; Status: Ongoing; Reported quantity: 1335 units; Lot: a) I68905, UDI/DI 00653160350944 (case) 10653160350941 (each), Lot Numbers: 2023012590, 2023042590; b) MNS11505, UDI/DI 00653160320213 (case) 10653160320210 (each), Lot Numbers: 2023022490, 2023071390; c) MNS11545, UDI/DI 00653160320305 (case) 10653160320302 (each), Lot Numbers: 2023021390; d) SUT21355, UDI/DI 00653160329223 (case) 10653160329220 (each), Lot Numbers: 2023060790. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- MEDLINE INDUSTRIES, LP - Northfield
- Distributor
- Worldwide distribution - US Nationwide and the countries of Panama, Canada.
- Lot
- a) I68905, UDI/DI 00653160350944 (case) 10653160350941 (each), Lot Numbers: 2023012590, 2023042590; b) MNS11505, UDI/DI 00653160320213 (case) 10653160320210 (each), Lot Numbers: 2023022490, 2023071390; c) MNS11545, UDI/DI 00653160320305 (case) 10653160320302 (each), Lot Numbers: 2023021390; d) SUT21355, UDI/DI 00653160329223 (case) 10653160329220 (each), Lot Numbers: 2023060790
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACURAD Alcohol Prep Pads, Sterile, Medium, 2-Ply, Contents: 5 boxes per Carton, 30 Boxes per Case, Single UseSeptember 3, 2025
- FDAReliOn, Sterile Alcohol Swabs, Skin Cleanser, 200 Swabs, Distributed by, Walmart, Inc., Bentonville, AR 72716September 3, 2025
- FDAGood Neighbor Pharmacy, Alcohol Prep Pads, Distributed By AmeriSource Bergen, 1 West First Avenue,September 3, 2025
- FDAMedline Kits containing BD SureStep Foley Tray Systems: SKU/Description: CDS984076J GYN ROBOTIC DYNJG001004ASeptember 3, 2025
- FDACenturion manual surgical kits labeled as: a) ST. VIRTUS SPLINTER FORCEP 6" (SF8246), Product Code 66580; b)April 10, 2024
- FDACenturion manual surgical kits labeled as: a) STERILE 4-1/2" S/S SCISSOR (SS5031), Product Code 65145; b)April 10, 2024
Questions about this recall
Was Centurion manual surgical kits labeled as: a) STERILE ALLIS TISS FCP 10", Product Code I68905; b) VAGINAL REPAIR KIT, Product Code MNS11505; c) PTCA ADD-ON KIT, Product Code MNS11545; d) INSTRUMENT SET, Product Code SUT21355 recalled?
Yes. FDA published recall Z-1443-2024 on April 10, 2024. Reported hazard: Product may have a weak seal which may result in a breach of sterility of the contents if the seal fails. The weak seal may not be detectable by all users.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit