Back to search
FDAFDA · Food and drugs·Recall number Z-1452-2013

Part 510031 Disposable Single Infusion Pump, quantity Includes... TOTAL VOLUME: 150mL, 2 mL/hr continuous Delivery Time : App. 72 hrs when filled to capacity Sterile EO... found in the following GoPump Kits: GoPump Kit 510042, w/ Epidural Catheter GoPump Kit 510042 - BP, w/ Epidural Catheter, BIOPATCH GoPump Kit 510110, w/2.5" Fenestrated Catheter GoPumpKit 510110-BP, w/2.5" Fenestrated Catheter,BioPatch GoPumpKit 510350 w/ Epidural Catheter GoPump Kit 510350-BP w/ Epidural Catheter, BIOPATCH GoPumpKit 510556-BP , 2.5 catheter, BIOPATCH Symbios GOPump Rapid Recovery System is a local pain mangagement system that consits of a small balloon that is inflated with a local anesthetic medication.

Published June 12, 2013 · Recall initiated: May 10, 2013

Summary

FDA records medical device Class I recall Z-1452-2013, published June 12, 2013 and initiated May 10, 2013. Recalling firm: Symbios Medical Products, LLC; Status: Terminated; Reported quantity: 33, 543 in all sizes/ kits; Lot: Pump 510031, Lot 11-l 00106 to Lot 12-101133 Kit 510042, lots 11-100251 to 13-100128 Kit 510042-BP , lot 12-101300 Kit 510110, lot 11-100664 Kit 510110-BP, Lots 11-100838 to 12-101560 Kit 510350, Lots 11-100353 to 12-100233 Kit 510350-BP, Lots12-100429 to 13-100047 Kit 510556-BP Lots 11-100779 to 12-101552. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Symbios Medical Products, LLC
Distributor
US Nationwide distribution.
Lot
Pump 510031, Lot 11-l 00106 to Lot 12-101133 Kit 510042, lots 11-100251 to 13-100128 Kit 510042-BP , lot 12-101300 Kit 510110, lot 11-100664 Kit 510110-BP, Lots 11-100838 to 12-101560 Kit 510350, Lots 11-100353 to 12-100233 Kit 510350-BP, Lots12-100429 to 13-100047 Kit 510556-BP Lots 11-100779 to 12-101552
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Part 510031 Disposable Single Infusion Pump, quantity Includes... TOTAL VOLUME: 150mL, 2 mL/hr continuous Delivery Time : App. 72 hrs when filled to capacity Sterile EO... found in the following GoPump Kits: GoPump Kit 510042, w/ Epidural Catheter GoPump Kit 510042 - BP, w/ Epidural Catheter, BIOPATCH GoPump Kit 510110, w/2.5" Fenestrated Catheter GoPumpKit 510110-BP, w/2.5" Fenestrated Catheter,BioPatch GoPumpKit 510350 w/ Epidural Catheter GoPump Kit 510350-BP w/ Epidural Catheter, BIOPATCH GoPumpKit 510556-BP , 2.5 catheter, BIOPATCH Symbios GOPump Rapid Recovery System is a local pain mangagement system that consits of a small balloon that is inflated with a local anesthetic medication. recalled?

Yes. FDA published recall Z-1452-2013 on June 12, 2013. Reported hazard: Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

What should I do if I have this product?

Voluntary: Firm initiated · Letter