Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Published April 10, 2024 · Recall initiated: February 14, 2024
Summary
FDA records medical device Class II recall Z-1457-2024, published April 10, 2024 and initiated February 14, 2024. Recalling firm: Covidien, LP; Status: Ongoing; Reported quantity: 30 Units; Lot: Model Number: SIGSDS30CTVT; UDI/DI: 10884521741881; Lot number: N3L1909UY. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Covidien, LP
- Distributor
- Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
- Lot
- Model Number: SIGSDS30CTVT; UDI/DI: 10884521741881; Lot number: N3L1909UY;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAV-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1August 19, 2026
- FDASurgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GXAugust 19, 2026
- FDAV-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436August 19, 2026
- FDADermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31August 19, 2026
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total KneeApril 10, 2024
- FDACovidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVTApril 10, 2024
Questions about this recall
Was Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT recalled?
Yes. FDA published recall Z-1457-2024 on April 10, 2024. Reported hazard: Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
What should I do if I have this product?
Voluntary: Firm initiated · Letter