Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration.
Published April 25, 2018 · Recall initiated: March 22, 2018
Summary
FDA records medical device Class II recall Z-1461-2018, published April 25, 2018 and initiated March 22, 2018. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 2,267; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Worldwide Distribution -- USA, Argentina, Bolivia, Brazil, Canada, Colombia, Guatemala, Mexico, Panama, Peru, Uruguay, and Venezuela.
- Lot
- All lots
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDAIbuprofen Tablets, USP, 200 mg, a) 50 tablet (NDC 53943-291-15), b) 100 tablet (NDC 53943-291-12), c) 500May 2, 2018
- FDAJAMSHIDI NEEDLE BIOPSY [15G X 100MM DISP, 16G X 100MM DISP, 17G X 100MM, 16G X 70MM DISP]; STERILE;April 25, 2018
Questions about this recall
Was Cook Vacuum Pump ( Cook Aspiration Unit) K-MAR-5200-US, GPN G49275. Intended for the aspiration of body fluids and cells, in particular oocyte aspiration. recalled?
Yes. FDA published recall Z-1461-2018 on April 25, 2018. Reported hazard: Pump was found to be constructed from crimp connectors that do not meet the fixation wiring requirements and can cause failure of the device to operate or electric shock or burn to the user.
What should I do if I have this product?
Voluntary: Firm initiated · Letter