Back to search
FDAFDA · Food and drugs·Recall number Z-1477-2021

Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R. Stopcocks and manifolds used to control the direction of IV fluid flow.

Published April 28, 2021 · Recall initiated: March 31, 2021

Summary

FDA records medical device Class II recall Z-1477-2021, published April 28, 2021 and initiated March 31, 2021. Recalling firm: Smiths Medical ASD Inc.; Status: Terminated; Reported quantity: 3825 units; Lot: a. Model Number MX1431MR, Lot Numbers: 3995544 b. Model Number MX1431MRL, Lot Numbers: 3995545, 4011838, 4027758, 4033501, 4044959 c. Model Number MX4331R, Lot Numbers: 3971467, 3987918. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
Distribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.
Lot
a. Model Number MX1431MR, Lot Numbers: 3995544 b. Model Number MX1431MRL, Lot Numbers: 3995545, 4011838, 4027758, 4033501, 4044959 c. Model Number MX4331R, Lot Numbers: 3971467, 3987918
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Stopcock rotator (Component): a. 1050 STOPCOCK ROTATOR, OFF HANDLE, Model Number MX1431MR b. 1050 STOPCOCK LEFT ROTATOR, OFF HANDLE, Model Number MX1431MRL c. 700 PSI STPCK W/ ROTATOR, Model Number MX4331R. Stopcocks and manifolds used to control the direction of IV fluid flow. recalled?

Yes. FDA published recall Z-1477-2021 on April 28, 2021. Reported hazard: Specific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail