SIGMA SPECTRUM Volumetric Infusion Pump with Master Drug Library Product Usage: Sigma Spectrum Volumetric Infusion Pumps with Master Drug Library are intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy.
Published April 30, 2014 · Recall initiated: February 7, 2014
Summary
FDA records medical device Class I recall Z-1484-2014, published April 30, 2014 and initiated February 7, 2014. Recalling firm: Baxter Healthcare Corp.; Status: Terminated; Reported quantity: 277,096 units (revised); Lot: Product Codes: 35700BAX, 35700ABB. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare Corporation has issued an Urgent Device Correction for the SIGMA SPECTRUM Volumetric Infusion Pump with Master Drug Library due to repeated System Error 322 occurrences.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corp.
- Distributor
- Worldwide Distribution - USA Nationwide including PR and the country of Canada.
- Lot
- Product Codes: 35700BAX, 35700ABB
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGAMBRO Capillary Dialyzer, Polyflux(R) Revaclear MAX, STERILE, Product code 110634 Product Usage: TheNovember 16, 2016
- FDASynovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, productAugust 24, 2016
- FDA0.9% Sodium Chloride Injection, USP, 250 mL VIAFLEX Plastic Container Bag, Rx Only, Manufactured by BaxterJune 15, 2016
- FDADextrose Injection,USP,70% 2000 mL VIAFLEX Plastic Container Bag, Pharmacy Bulk Package, Not For DirectJune 15, 2016
- FDARaySearch Ray Station Radiation Therapy Treatment Planning System, Stand-alone software. Versions 2.0, 2.5,April 30, 2014
- FDAMerit Laureate Hydrophilic Guide Wire, Catalog No. LWSTFS35260EX, Straight Tip Stiff Shaft, 0.035" (0.89mm),April 30, 2014
Questions about this recall
Was SIGMA SPECTRUM Volumetric Infusion Pump with Master Drug Library Product Usage: Sigma Spectrum Volumetric Infusion Pumps with Master Drug Library are intended to be used for the controlled administration of intravenous fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixtures of required patient therapy. recalled?
Yes. FDA published recall Z-1484-2014 on April 30, 2014. Reported hazard: Baxter Healthcare Corporation has issued an Urgent Device Correction for the SIGMA SPECTRUM Volumetric Infusion Pump with Master Drug Library due to repeated System Error 322 occurrences.
What should I do if I have this product?
Voluntary: Firm initiated · Letter