FDAFDA · Food and drugs·Recall number Z-1497-2013

Stryker Instruments 0400-820-000S3 T4 Toga w/Inside Shoulder Ties, Sterile, O; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.

Published June 12, 2013 · Recall initiated: May 23, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Stryker Instruments Div. of Stryker Corporation
Distributor
USA Nationwide Distribution
Lot
1007145, 1101004, 1107487
Risk classification
Class II
Area (FDA)
device
Country
United States