LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.
Published March 25, 2020 · Recall initiated: February 24, 2020
Summary
FDA records medical device Class II recall Z-1498-2020, published March 25, 2020 and initiated February 24, 2020. Recalling firm: LivaNova USA Inc.; Status: Terminated; Reported quantity: Total Worldwide=9309 systems (2713 system in U.S.); Lot: Device Listing: E572259 Product Code Product Description Affected Serial Number range 16-02-80 Heater-Cooler 3T, 230V 16S10027 16S16874 16-02-81 Heater-Cooler 3T, 240V 16S10743 16S11708 16-02-82 Heater-Cooler 3T, 208V 16S10772 16S16840 16-02-83 Heater-Cooler 3T, 127V 16S11455 16S16509 16-02-85 Heater-Cooler 3T, 120V 16S10958 16S16847 16-02-95 Heater-Cooler 3T, 200V 16S12004 16S16818. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- LivaNova USA Inc.
- Distributor
- U.S. Nationwide distribution including Puerto Rico. This matter was triggered by the clearance of K191402, thus only the United States, including Puerto Rico, is affected. O.U.S. - none
- Lot
- Device Listing: E572259 Product Code Product Description Affected Serial Number range 16-02-80 Heater-Cooler 3T, 230V 16S10027 16S16874 16-02-81 Heater-Cooler 3T, 240V 16S10743 16S11708 16-02-82 Heater-Cooler 3T, 208V 16S10772 16S16840 16-02-83 Heater-Cooler 3T, 127V 16S11455 16S16509 16-02-85 Heater-Cooler 3T, 120V 16S10958 16S16847 16-02-95 Heater-Cooler 3T, 200V 16S12004 16S16818
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASterile Disposable Connectors ASY 1/2 CONN W/LL STRLE 24 REF/ GTIN for Insert Label:050608000/ 00803622123382July 8, 2020
- FDAVein Irrigation Cannula with blunt tip. ASY VIC, STERILE 25/CS REF/ GTIN for Insert Label: VIC/ 00803622133107July 8, 2020
- FDASterile Disposable Connectors ASY RDC Y 1/2X3/8X3/8 STR 24 REF/ GTIN for Insert Label:050529000/July 8, 2020
- FDASterile Disposable Connectors ASY STR CONN 1/4 STRLE 24. REF/ GTIN for Insert Label:050504000/ 00803622123221July 8, 2020
- FDAArtis Q.zen Biplane, Model No. 10848355 Artis Q.zen Ceiling, Model No. 10848354 Artis Q.zen Floor, Model No.March 25, 2020
- FDAAGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage:March 25, 2020
Questions about this recall
Was LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less. recalled?
Yes. FDA published recall Z-1498-2020 on March 25, 2020. Reported hazard: The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
What should I do if I have this product?
Voluntary: Firm initiated · Letter