Back to search
FDAFDA · Food and drugs·Recall number Z-1515-2014

The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0 mm SS Sleeve Only; Part Number 01-0022-S 8, 7 hole, SS compression plate; Part Number 02-2007-S 10, 9 hole, SS compression plate; Part Number 02-2009-S Product Usage: The Osteo-Clage Stainless Steel Bone Plate is an implantable straight, rigid stainless steel plate. It is used in conjunction with 4.5mm Cortical bone screws to provide compression across bone fractures. The Osteo-Clage Cerclage Wire Crimp sleeve is used in conjunction with the bone plate in cerclage fixation procedures

Published May 7, 2014 · Recall initiated: March 26, 2014

Summary

FDA records medical device Class II recall Z-1515-2014, published May 7, 2014 and initiated March 26, 2014. Recalling firm: Acumed LLC; Status: Terminated; Reported quantity: 141 units in the US; 294 units outside the US.; Lot: 2.0mm x 80mm SS cable/sleeve: Lot # 222078, 232175, 245032, 254368, 261903, 270659, 284155, 297281. 2.0mm SS sleeve only: Lot # 247553, 267469, 280871, 313261. 8, 7 hole SS compression plate: Lot # 300381. 10, 9 hole SS compression plate: Lot # 299641. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Manufacturing of these devices with a grade of stainless steel that is not within specifications.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Acumed LLC
Distributor
US Nationwide Distribution - in the states of CA, CO, FL, GA, LA, MI, MN, MO, OH, OR, PA, PR, TX, UT, VA, WA and WY. Product was distributed outside the US to: Australia, China, Figi, France, Great Britain, Italy, Japan, Korea, Maylasia, Sweden, South Africa, Spain.
Lot
2.0mm x 80mm SS cable/sleeve: Lot # 222078, 232175, 245032, 254368, 261903, 270659, 284155, 297281. 2.0mm SS sleeve only: Lot # 247553, 267469, 280871, 313261. 8, 7 hole SS compression plate: Lot # 300381. 10, 9 hole SS compression plate: Lot # 299641.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was The Osteo-Clage System consists of stainless steel cable/sleeves and compression plates. . 2.0 mm x 850mm SS Cable/Sleeve; Part Number 01-0020-S 2.0 mm SS Sleeve Only; Part Number 01-0022-S 8, 7 hole, SS compression plate; Part Number 02-2007-S 10, 9 hole, SS compression plate; Part Number 02-2009-S Product Usage: The Osteo-Clage Stainless Steel Bone Plate is an implantable straight, rigid stainless steel plate. It is used in conjunction with 4.5mm Cortical bone screws to provide compression across bone fractures. The Osteo-Clage Cerclage Wire Crimp sleeve is used in conjunction with the bone plate in cerclage fixation procedures recalled?