Integra CUSA Excel Disposable Wrench. Catalog Number C5601. The CUSA Excel+ Sterile 23 kHz Torque Wrench is an accessory used with the CUSA Excel+ System ultrasonic surgical instrument. The torque wrench is used to properly secure the CUSA tip to the CUSA handpiece in the sterile field.
Published May 7, 2014 · Recall initiated: March 10, 2014
Summary
FDA records medical device Class II recall Z-1522-2014, published May 7, 2014 and initiated March 10, 2014. Recalling firm: Integra LifeSciences Corp.; Status: Terminated; Reported quantity: 3,984 devices packaged as 6 devices / box (664 boxes); Lot: Product/Catalog Number C5601; Lot Numbers 1132424, 1132883, 1132884, 1133453, 1133454, 1133455, 1134067, 1134675, 1134873 and 1135014. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Integra LifeSciences Corp.
- Distributor
- Worldwide Distribution - USA (nationwide) including the countries of Australia, Canada, and Switzerland.
- Lot
- Product/Catalog Number C5601; Lot Numbers 1132424, 1132883, 1132884, 1133453, 1133454, 1133455, 1134067, 1134675, 1134873 and 1135014
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: MediHoney® Gel Product Name: MediHoney® Wound & Burn Dressing Gel Model/Catalog Number: 31815February 19, 2025
- FDASURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicatedNovember 6, 2024
- FDASURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for theNovember 6, 2024
- FDASURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated forNovember 6, 2024
- FDAFisherbrand Sterile Swabs, Calcium Alginate Fiber Tipped Aluminum Applicator Swab, .025 and .035 in Diameter,May 7, 2014
- FDAMedtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713),May 7, 2014
Questions about this recall
Was Integra CUSA Excel Disposable Wrench. Catalog Number C5601. The CUSA Excel+ Sterile 23 kHz Torque Wrench is an accessory used with the CUSA Excel+ System ultrasonic surgical instrument. The torque wrench is used to properly secure the CUSA tip to the CUSA handpiece in the sterile field. recalled?
Yes. FDA published recall Z-1522-2014 on May 7, 2014. Reported hazard: Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert.
What should I do if I have this product?
Voluntary: Firm initiated · Letter