Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), consisting of the Monitor and test strip, is used for quantitative determination of international normalized ratio (INR) in fresh, capillary whole blood to monitor the effect of warfarin therapy on clotting time by health care professionals. (in vitro diagnostic use).
Published May 14, 2014 · Recall initiated: April 16, 2014
Summary
FDA records medical device Class I recall Z-1546-2014, published May 14, 2014 and initiated April 16, 2014. Recalling firm: Alere San Diego, Inc.; Status: Terminated; Reported quantity: Estimated 99,795 Total units; Lot: Lot/Unit Codes: Applicable to all Alere INRatio2 PT/INR Professional Test Strips lots packaged from 8/22/13 through 3/28/14, including the following: 324810, 324811, 324812, 327543, 327544, 327549, 328109, 328705, 328707, 330238, 330852, 331340, 331916, 331922, 332841, 332842, 332843, 332959, 333362, 333654, 333655, 333864, 333865, 334572, 334575, 335552, 335553, 335796, 335797, 336426, 336668, 337350, 338965, 338966, 339257, 339707, 327856V Expiration date, or Expected shelf life: range from 4/30/14 to 1/31/15. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Alere San Diego, Inc.
- Distributor
- US Nationwide Distribution
- Lot
- Lot/Unit Codes: Applicable to all Alere INRatio2 PT/INR Professional Test Strips lots packaged from 8/22/13 through 3/28/14, including the following: 324810, 324811, 324812, 327543, 327544, 327549, 328109, 328705, 328707, 330238, 330852, 331340, 331916, 331922, 332841, 332842, 332843, 332959, 333362, 333654, 333655, 333864, 333865, 334572, 334575, 335552, 335553, 335796, 335797, 336426, 336668, 337350, 338965, 338966, 339257, 339707, 327856V Expiration date, or Expected shelf life: range from 4/30/14 to 1/31/15
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine TestOctober 7, 2026
- FDAFisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + UrineOctober 7, 2026
- FDAMcKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine,October 7, 2026
- FDAFisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman CoulterOctober 7, 2026
- FDACLINITEK Status+ Analyzer (portable urine chemistry analyzer). Catalog number 10490946.May 14, 2014
- FDAGiraffe® Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable toMay 14, 2014
Questions about this recall
Was Alere INRatio2 PT/INR Professional Test Strips, Model Number: 99008G2. The Alere INRatio2 PT/INR Monitoring System (Professional Use), consisting of the Monitor and test strip, is used for quantitative determination of international normalized ratio (INR) in fresh, capillary whole blood to monitor the effect of warfarin therapy on clotting time by health care professionals. (in vitro diagnostic use). recalled?
Yes. FDA published recall Z-1546-2014 on May 14, 2014. Reported hazard: Test Strips may report an inaccurately low INR result. Several patients had a therapeutic or near-therapeutic INR with the Alere INRatio2 PT/INR Professional Test Strip but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.
What should I do if I have this product?
Voluntary: Firm initiated · Letter