Panda® Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).
Published May 14, 2014 · Recall initiated: April 2, 2014
Summary
FDA records medical device Class II recall Z-1549-2014, published May 14, 2014 and initiated April 2, 2014. Recalling firm: GE Healthcare; Status: Terminated; Reported quantity: 21 units; Lot: HDJS51142 HDJS52094 HDJS52095 HDJS52153 HDJS52154 HDJS52155 HDJS52156 HDJS52223 HDJS52224 HDJS52517 HDJS52518 HDJS52519 HDJS52520 HDJS52521 HDJS52522 HDJS52974 HDJS52975 HDJS52976 HDJS52977 HDJS52978 HDJS52979. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Healthcare
- Distributor
- Worldwide Distribution.
- Lot
- HDJS51142 HDJS52094 HDJS52095 HDJS52153 HDJS52154 HDJS52155 HDJS52156 HDJS52223 HDJS52224 HDJS52517 HDJS52518 HDJS52519 HDJS52520 HDJS52521 HDJS52522 HDJS52974 HDJS52975 HDJS52976 HDJS52977 HDJS52978 HDJS52979
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACentricity Enterprise Archive V2, V3, and V4, software product for receiving, archiving, and sendingSeptember 23, 2026
- FDAWorkflow Core Services V8, intended to provide set of services for storage and retrieval of patient data,September 23, 2026
- FDACentricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review,September 23, 2026
- FDACore Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC)September 23, 2026
- FDAPanda® iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unableMay 14, 2014
- FDAPanda® Wall-Mount Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who areMay 14, 2014
Questions about this recall
Was Panda® Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor). recalled?
Yes. FDA published recall Z-1549-2014 on May 14, 2014. Reported hazard: Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
What should I do if I have this product?
Voluntary: Firm initiated · Letter