Fixt Suture Bonded Plus Violet Braided Polyglycolic Acid Coated Suture, Synthetic Absorbable Surgical Suture, Sterile, Rx Only, Teleflex Medical. Indicated for use in soft tissue approximation.
Published May 14, 2014 · Recall initiated: March 11, 2014
Summary
FDA records medical device Class II recall Z-1558-2014, published May 14, 2014 and initiated March 11, 2014. Recalling firm: Teleflex Medical; Status: Terminated; Reported quantity: 32,271 ea.; Lot: Product Code BON100 - Lot 02H1302839, 02J1101705, 02D1101137, 02F1103013, Product Code EP4049N - Lot 02A1003137. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Teleflex Medical
- Distributor
- Worldwide Distribution -- USA, including the states of AL, CA, FL, GA, IL, IN, KS, LA, ME, MA, MI, MN, MS, MO, NV, NJ, NM, NY, OH, PA, RI, SC, TN, TX, UT, VA, WV, and WI; and countries of Canada, Germany, and Ireland.
- Lot
- Product Code BON100 - Lot 02H1302839, 02J1101705, 02D1101137, 02F1103013, Product Code EP4049N - Lot 02A1003137
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm,September 18, 2019
- FDAHudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REFSeptember 18, 2019
- FDAHudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm,June 26, 2019
- FDARUSCH LASERTUBE (Rubber), Order No. : 102004-000050, REF 102004, SIZE I.D. mm 5,0June 5, 2019
- FDATeleflex Medical, Research Triangle Park, NC, Polypropylene, Rx Only, Sterile, Distributed by BostonMay 14, 2014
- FDABard Fixt Suture Braided Polyester, Rx Only, Teleflex Medical, Research Triangle Park, NC USA Assembled inMay 14, 2014
Questions about this recall
Was Fixt Suture Bonded Plus Violet Braided Polyglycolic Acid Coated Suture, Synthetic Absorbable Surgical Suture, Sterile, Rx Only, Teleflex Medical. Indicated for use in soft tissue approximation. recalled?
Yes. FDA published recall Z-1558-2014 on May 14, 2014. Reported hazard: Product does not meet minimum needle attachment strength requirements; thus, the needle can become detached from the suture during use.
What should I do if I have this product?
Voluntary: Firm initiated · Letter