Back to search
FDAFDA · Food and drugs·Recall number Z-1560-2020

Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Published April 1, 2020 · Recall initiated: February 3, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.

What to do

Voluntary: Firm initiated · Other

Recall details

Manufacturer
Dexcom Inc
Distributor
World-wide Distribution
Lot
Device Listing Number: D189069 Model: STS-GL-041 Description: G4/G5 Sensor Kit 4 Pack GUDID Registered GTIN: 30386270000010 Lot Number: All Model: STS-RX-041 Description: G4/G5 Sensor Kit 4 Pack Retail GUDID Registered GTIN: 20386270000013 Lot Number: All Model: STS-MC-004 Description: G4/G5 Sensor Kit 4 Pack Medicare GUDID Registered GTIN: 00386270000637 Lot Number: All
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card