TRUE Dilatation Balloon Valvuloplasty Catheter, REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes: 20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm; Intended to be used for balloon aortic valvuloplasty.
Published July 3, 2013 · Recall initiated: March 18, 2013
Summary
FDA records medical device Class II recall Z-1566-2013, published July 3, 2013 and initiated March 18, 2013. Recalling firm: Loma Vista Medical; Status: Terminated; Reported quantity: 2379 units; Lot: All lot numbers lower than PLN00749 are subject to this recall. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Loma Vista Medical
- Distributor
- Worldwide Distribution-US Distribution in New Jersey and the countries of Germany, Switzerland, Italy, Norway, The Netherlands, Spain, Portugal, Ireland, and Austria.
- Lot
- All lot numbers lower than PLN00749 are subject to this recall.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABaxter, MiniCap with Povidone-Iodine solution, 5C4466P, Baxter Healthcare Corporation. For dialysis. IntendedJuly 3, 2013
- FDACARESCAPE Monitor B850 Product Usage: The CARESCAPE TM Monitor B850 is a multi-parameter high acuity patientJuly 3, 2013
- FDAMedtronic, Launcher Guiding Catheter, Model # LA8EBU45SH, Lot # 0006722866, 8F and Model # LA6IMAD, Lot #July 3, 2013
- FDACMC Saddle Product Usage: The Ascension CMC is intended to resurface the proximal end of the first metacarpalJuly 3, 2013
- FDAThermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured andJuly 3, 2013
- FDASarns " TCM II Temperature Control Module The Sarns" TCM II Temperature Control Module is a source ofJuly 3, 2013
Questions about this recall
Was TRUE Dilatation Balloon Valvuloplasty Catheter, REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes: 20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm; Intended to be used for balloon aortic valvuloplasty. recalled?
Yes. FDA published recall Z-1566-2013 on July 3, 2013. Reported hazard: Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.
What should I do if I have this product?
Voluntary: Firm initiated · Letter