FDAFDA · Food and drugs·Recall number Z-1566-2013

TRUE Dilatation Balloon Valvuloplasty Catheter, REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes: 20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm; Intended to be used for balloon aortic valvuloplasty.

Published July 3, 2013 · Recall initiated: March 18, 2013

Summary

FDA records medical device Class II recall Z-1566-2013, published July 3, 2013 and initiated March 18, 2013. Recalling firm: Loma Vista Medical; Status: Terminated; Reported quantity: 2379 units; Lot: All lot numbers lower than PLN00749 are subject to this recall. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Loma Vista Medical
Distributor
Worldwide Distribution-US Distribution in New Jersey and the countries of Germany, Switzerland, Italy, Norway, The Netherlands, Spain, Portugal, Ireland, and Austria.
Lot
All lot numbers lower than PLN00749 are subject to this recall.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was TRUE Dilatation Balloon Valvuloplasty Catheter, REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes: 20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm; Intended to be used for balloon aortic valvuloplasty. recalled?

Yes. FDA published recall Z-1566-2013 on July 3, 2013. Reported hazard: Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds. The balloon normally deflates in less than 5 seconds. The device performance specification for deflation time is 10 seconds maximum.

What should I do if I have this product?

Voluntary: Firm initiated · Letter