LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the removal of emboli and thrombi during embolectomy and/or thrombectomy. It can also be used for catheter placement over a guidewire, vessel occlusion, fluid infusion and/or aspiration.

Published April 8, 2020 · Recall initiated: March 23, 2020

Summary

FDA records medical device Class I recall Z-1567-2020, published April 8, 2020 and initiated March 23, 2020. Recalling firm: LeMaitre Vascular, Inc.; Status: Ongoing; Reported quantity: 43315; Lot: OTW2897® OTW3273® OTW3333® OTW3379® OTW3418® OTW3543® OTW3564® OTW3620® OTW3700® OTW3740® OTW3758® OTW3805® OTW3865® OTW3879® OTW3892® OTW3966® OTW4009® OTW4010® OTW4017® OTW4018® OTW4024® OTW4060® OTW4063® OTW4064® OTW4065® OTW4117® OTW4201® OTW4238® OTW4239® OTW2957® OTW3001® OTW3266® OTW3395® OTW3565® OTW3637® OTW3638® OTW3658® OTW3665® OTW3670® OTW3671® OTW3682® OTW3683® OTW3694® OTW3695® OTW3701® OTW3708® OTW3727® OTW3728® OTW3747® OTW3769® OTW3770® OTW3775® OTW3776® OTW3779® OTW3780® OTW3810® OTW3820® OTW3821® OTW3853® OTW3854® OTW3880® OTW3881® OTW3886® OTW3893® OTW3898® OTW3899® OTW3915® OTW3916® OTW3924® OTW3930® OTW3957® OTW3958® OTW3981® OTW3982® OTW4011® OTW4025® OTW4026® OTW4030® OTW4031® OTW4052® OTW4053® OTW4054® OTW4066® OTW4067® OTW4068® OTW4073® OTW4074® OTW4075® OTW4092® OTW4093® OTW4098® OTW4099® OTW4105® OTW4106® OTW4107® OTW4119® OTW4120® OTW4157® OTW4158® OTW4188® OTW4197® OTW4220® OTW4221® OTW4222® OTW4229® OTW4230® OTW4247® OTW4249® OTW4254® OTW4255® OTW4258® OTW4259® OTW4260® OTW4261® OTW4263® OTW4291® OTW4294® OTW4339® OTW4341® OTW4363® OTW4364® OTW2942® OTW2959® OTW2987® OTW3207® OTW3292® OTW3303® OTW3312® OTW3322® OTW3323® OTW3351® OTW3367® OTW3438A® OTW3511® OTW3635® OTW3636® OTW3660® OTW3661® OTW3667® OTW3688® OTW3689® OTW3706® OTW3707® OTW3711® OTW3717® OTW3718® OTW3734® OTW3750® OTW3783® OTW3784® OTW3794® OTW3795® OTW3827® OTW3828® OTW3851® OTW3852® OTW3862® OTW3863® OTW3882® OTW3896® OTW3897® OTW3907® OTW3908® OTW3931® OTW3936® OTW3944® OTW3945® OTW3962® OTW3963® OTW3978® OTW3979® OTW3995® OTW3996® OTW4033® OTW4034® OTW4057® OTW4058® OTW4084® OTW4085® OTW4127® OTW4128® OTW4140® OTW4141® OTW4146® OTW4147® OTW4176® OTW4177® OTW4183® OTW4184® OTW4191® OTW4192® OTW4193® OTW4232® OTW4317® OTW4336® OTW2872® OTW3191® OTW3439® OTW3537® OTW3623® OTW3662® OTW3668® OTW3669® OTW3690® OTW3691® OTW3712® OTW3735® OTW3766® OTW3771® OTW3812® OTW3822® OTW3823® OTW3836® OTW3849® OTW3850® OTW3874® OTW3888® OTW3903® OTW3925® OTW3932® OTW3946® OTW3947® OTW3968® OTW3969® OTW3987® OTW3988® OTW4020® OTW4021® OTW4036® OTW4037® OTW4038® OTW4079® OTW4080® OTW4087® OTW4196® OTW4267® OTW4268® OTW4269® OTW4270® OTW4275® OTW4276® OTW4277® OTW4288® ***Added 3/19/20*** OTW2847® OTW2863® OTW2871® ***Updated 4/14/20*** The firm noted upon review that some of the original recall lots were originally relabeled with different lot numbers for Non-US customers. Those additional lot numbers are as follow: OTW3461 OTW3501 OTW3591 OTW3628 OTW3737 OTW3742 OTW3753 OTW3813 OTW3859 OTW3861 OTW4208 ***Updated 8/21/20*** Lots OTW4180, OTW4187, OTW4198, OTW4202, OTW4218, OTW4246. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
LeMaitre Vascular, Inc.
Distributor
Worldwide distribution: US (Nationwide distribution) and foreign countries of: Australia, Austria, Belgium, Bulgaria, Colombia, Costa Rica, Czech Republic, Dominican Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Georgia, Hungary, Ireland, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom, and Vietnam.
Lot
OTW2897® OTW3273® OTW3333® OTW3379® OTW3418® OTW3543® OTW3564® OTW3620® OTW3700® OTW3740® OTW3758® OTW3805® OTW3865® OTW3879® OTW3892® OTW3966® OTW4009® OTW4010® OTW4017® OTW4018® OTW4024® OTW4060® OTW4063® OTW4064® OTW4065® OTW4117® OTW4201® OTW4238® OTW4239® OTW2957® OTW3001® OTW3266® OTW3395® OTW3565® OTW3637® OTW3638® OTW3658® OTW3665® OTW3670® OTW3671® OTW3682® OTW3683® OTW3694® OTW3695® OTW3701® OTW3708® OTW3727® OTW3728® OTW3747® OTW3769® OTW3770® OTW3775® OTW3776® OTW3779® OTW3780® OTW3810® OTW3820® OTW3821® OTW3853® OTW3854® OTW3880® OTW3881® OTW3886® OTW3893® OTW3898® OTW3899® OTW3915® OTW3916® OTW3924® OTW3930® OTW3957® OTW3958® OTW3981® OTW3982® OTW4011® OTW4025® OTW4026® OTW4030® OTW4031® OTW4052® OTW4053® OTW4054® OTW4066® OTW4067® OTW4068® OTW4073® OTW4074® OTW4075® OTW4092® OTW4093® OTW4098® OTW4099® OTW4105® OTW4106® OTW4107® OTW4119® OTW4120® OTW4157® OTW4158® OTW4188® OTW4197® OTW4220® OTW4221® OTW4222® OTW4229® OTW4230® OTW4247® OTW4249® OTW4254® OTW4255® OTW4258® OTW4259® OTW4260® OTW4261® OTW4263® OTW4291® OTW4294® OTW4339® OTW4341® OTW4363® OTW4364® OTW2942® OTW2959® OTW2987® OTW3207® OTW3292® OTW3303® OTW3312® OTW3322® OTW3323® OTW3351® OTW3367® OTW3438A® OTW3511® OTW3635® OTW3636® OTW3660® OTW3661® OTW3667® OTW3688® OTW3689® OTW3706® OTW3707® OTW3711® OTW3717® OTW3718® OTW3734® OTW3750® OTW3783® OTW3784® OTW3794® OTW3795® OTW3827® OTW3828® OTW3851® OTW3852® OTW3862® OTW3863® OTW3882® OTW3896® OTW3897® OTW3907® OTW3908® OTW3931® OTW3936® OTW3944® OTW3945® OTW3962® OTW3963® OTW3978® OTW3979® OTW3995® OTW3996® OTW4033® OTW4034® OTW4057® OTW4058® OTW4084® OTW4085® OTW4127® OTW4128® OTW4140® OTW4141® OTW4146® OTW4147® OTW4176® OTW4177® OTW4183® OTW4184® OTW4191® OTW4192® OTW4193® OTW4232® OTW4317® OTW4336® OTW2872® OTW3191® OTW3439® OTW3537® OTW3623® OTW3662® OTW3668® OTW3669® OTW3690® OTW3691® OTW3712® OTW3735® OTW3766® OTW3771® OTW3812® OTW3822® OTW3823® OTW3836® OTW3849® OTW3850® OTW3874® OTW3888® OTW3903® OTW3925® OTW3932® OTW3946® OTW3947® OTW3968® OTW3969® OTW3987® OTW3988® OTW4020® OTW4021® OTW4036® OTW4037® OTW4038® OTW4079® OTW4080® OTW4087® OTW4196® OTW4267® OTW4268® OTW4269® OTW4270® OTW4275® OTW4276® OTW4277® OTW4288® ***Added 3/19/20*** OTW2847® OTW2863® OTW2871® ***Updated 4/14/20*** The firm noted upon review that some of the original recall lots were originally relabeled with different lot numbers for Non-US customers. Those additional lot numbers are as follow: OTW3461 OTW3501 OTW3591 OTW3628 OTW3737 OTW3742 OTW3753 OTW3813 OTW3859 OTW3861 OTW4208 ***Updated 8/21/20*** Lots OTW4180, OTW4187, OTW4198, OTW4202, OTW4218, OTW4246
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Over the Wire Embolectomy Catheter is indicated for use in the removal of emboli and thrombi during embolectomy and/or thrombectomy. It can also be used for catheter placement over a guidewire, vessel occlusion, fluid infusion and/or aspiration. recalled?

Yes. FDA published recall Z-1567-2020 on April 8, 2020. Reported hazard: There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation.

What should I do if I have this product?

Voluntary: Firm initiated · Letter