Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve to be used with Medtronic SynchroMed implantable drug infusion pumps. Contents of inner package are STERILE. The Medtronic Model 8578 Sutureless Pump Connector Revision Kit is used when a pump connector for an Indura 1P Model 8709 or Model 8709SC catheter is required. The catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Indura 1P Model 8709 or Model 8709SC catheter. The catheter connects to the pump with the Model 8578 sutureless pump connector at the catheter port.
Published July 3, 2013 · Recall initiated: June 3, 2013
Summary
FDA records medical device Class I recall Z-1573-2013, published July 3, 2013 and initiated June 3, 2013. Recalling firm: Medtronic Neuromodulation; Status: Terminated; Reported quantity: 115,722 total; Lot: Product having a Use By Date prior to 25 Aug 2014. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Neuromodulation
- Distributor
- Worldwide distribution: US (nationwide) and countries of: Aruba, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Lebanon, Lithuania, Luxembourg, Malta, Martinique, Morocco, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Venezuela.
- Lot
- Product having a Use By Date prior to 25 Aug 2014
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInterStim SureScan MRI Leads 978A1 & 978B1: LEAD 978A128 ISTM MICRO 28CM SSMRI TINED 978A128 LEAD 978A133 ISTMOctober 28, 2026
- FDAInterStim Percutaneous Extensions EXT 3560022 ISTM II PERC 3560022 EXT 3560030 ISTM MICRO PERC 3560030October 28, 2026
- FDAMedtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted,June 3, 2026
- FDAMedtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted,June 3, 2026
- FDAMAQUET GETINGE GROUP Arterial Cannula OD Used to introduce blood and sterile solutions into the systemicJuly 3, 2013
- FDAMedtronic Intrathecal Catheter Pump Segment Revision Kit, model 8596SC. Contents: 60-cm pump segment withJuly 3, 2013
Questions about this recall
Was Medtronic Sutureless Pump Connector Revision Kit, model 8578. Contents: catheter interface with attached sutureless pump connector, catheter, connector pin, and strain-relief sleeve to be used with Medtronic SynchroMed implantable drug infusion pumps. Contents of inner package are STERILE. The Medtronic Model 8578 Sutureless Pump Connector Revision Kit is used when a pump connector for an Indura 1P Model 8709 or Model 8709SC catheter is required. The catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space. The implanted infusion system components consist of a Medtronic pump and an Indura 1P Model 8709 or Model 8709SC catheter. The catheter connects to the pump with the Model 8578 sutureless pump connector at the catheter port. recalled?
Yes. FDA published recall Z-1573-2013 on July 3, 2013. Reported hazard: The Sutureless Connector (SC) Intrathecal Catheter connector has been redesigned to reduce the potential for occlusion at the catheter to pump interface. Medtronic is removing the unused products from the market that were manufactured with the previous design, and recommend the previous design no longer be used due to greater potential for misalignment and subsequent occlusion.
What should I do if I have this product?
Voluntary: Firm initiated · Letter