Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143
Published April 24, 2024 · Recall initiated: March 21, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- US Nationwide distribution.
- Lot
- Model/UDI-DI: Artis zee floor- 4056869010045 Artis zee ceiling- 4056869010052 Artis zee multi-purpose- 4056869010076 Artis zee biplane- 4056869010069 Artis zee floor MN- 4056869010090 Artis zee biplane MN- 4056869010106 Artis zeego- 4056869010083 Artis Q floor- 4056869009971 Artis Q ceiling- 4056869009988 Artis Q biplane- 4056869009995 Artis Q zeego- 4056869010007 Artis Q.zen floor- 4056869010014 Artis Q.zen ceiling- 4056869010021 Artis Q.zen biplane- 4056869010038
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.