Alinity m System, Part Number: 08N53-002
Published May 1, 2024 · Recall initiated: March 20, 2024
Summary
FDA records medical device Class II recall Z-1602-2024, published May 1, 2024 and initiated March 20, 2024. Recalling firm: Abbott Molecular, Inc.; Status: Ongoing; Reported quantity: 1,012 units; Lot: Part Number: 08N53-002; UD/DI: 00884999048034. Not Serial Number specific. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Molecular, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CAMEROON, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KENYA, LATVIA, LUXEMBOURG, MALAWI, MALAYSIA, MEXICO, MONTENEGRO, MOROCCO, MOZAMBIQUE, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, PAKISTAN, PANAMA, PARAGUAY, POLAND, PORTUGAL, ROMANIA, RUSSIA (CIS), SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TURKEY, UGANDA, UNITED KINGDOM, VIETNAM, ZIMBABWE.
- Lot
- Part Number: 08N53-002; UD/DI: 00884999048034. Not Serial Number specific
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included inApril 30, 2025
- FDAAbbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REFNovember 6, 2024
- FDAAbbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REFNovember 6, 2024
- FDAACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running softwareMay 1, 2024
- FDAIBA Proton Therapy System - PROTEUS 235, PTS-10 versions, PTS-11 versions, and PTS-12 versions.May 1, 2024
- FDAIn-vitro diagnostic kits: (1) Abbott Vysis MDM2 / CEP 12 FISH Probe Kit, REF 01N15-010; (2) Abbott Vysis LSIMarch 6, 2024
Questions about this recall
Was Alinity m System, Part Number: 08N53-002 recalled?
Yes. FDA published recall Z-1602-2024 on May 1, 2024. Reported hazard: The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
What should I do if I have this product?
Voluntary: Firm initiated · Letter