Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.
Published May 1, 2024 · Recall initiated: February 15, 2024
Summary
FDA records medical device Class II recall Z-1605-2024, published May 1, 2024 and initiated February 15, 2024. Recalling firm: SEASPINE ORTHOPEDICS CORPORATION; Status: Ongoing; Reported quantity: 1579 Screw Heads; Lot: REF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- SEASPINE ORTHOPEDICS CORPORATION
- Distributor
- US Nationwide distribution in the states of MI, GA, VA, IN, FL.
- Lot
- REF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASeaSpine Mariner Deformity Product Name: Extended Polyaxial Head, Attached Gimbal, 12mm REF: MD1-300010-12September 30, 2026
- FDASeaSpine Polyaxial Head REF Number: 41-3010July 29, 2026
- FDABrand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/CatalogJanuary 1, 2025
- FDABrand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 ProductJuly 24, 2024
- FDARedux Electrolyte Creme, Product Ref. 66-04. Highly conductive electrolyte creme.May 1, 2024
- FDAVarious Alcon Custom Paks Surgical Procedure Pack containing Oasis 27Ga Chang Hydrodissection Cannulas. TheseMay 1, 2024
Questions about this recall
Was Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems. recalled?
Yes. FDA published recall Z-1605-2024 on May 1, 2024. Reported hazard: Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
What should I do if I have this product?
Voluntary: Firm initiated · Letter