NexGen CR-Flex Femoral Component, Size E and Size E Minus (E-). Component of the NexGen CR-Flex Porous Femur System.
Published May 1, 2024 · Recall initiated: March 28, 2024
Summary
FDA records medical device Class II recall Z-1608-2024, published May 1, 2024 and initiated March 28, 2024. Recalling firm: Zimmer, Inc.; Status: Ongoing; Reported quantity: 5,546 units; Lot: Item No. 00595201501, 00595201502, 00595201505, 00595201506; UDI:(01)00889024207806(17)280731(10)64115713, (01)00889024207806(17)280930(10)64175075, (01)00889024207806(17)280930(10)64175076, (01)00889024207790(17)280930(10)64175073, (01)00889024207806(17)280930(10)64138317, (01)00889024207790(17)281031(10)64175074, (01)00889024207790(17)281031(10)64194991, (01)00889024207790(17)281031(10)64202116, (01)00889024207806(17)281031(10)64178998, (01)00889024207813(17)281031(10)64179054, (01)00889024207790(17)281031(10)64202117, (01)00889024207813(17)281031(10)64179043, (01)00889024207813(17)281031(10)64194989, (01)00889024207790(17)281031(10)64178996, (01)00889024207790(17)281031(10)64178997, (01)00889024207790(17)281031(10)64179023, (01)00889024207820(17)281031(10)64178999, (01)00889024207813(17)281031(10)64179043R, (01)00889024207813(17)281031(10)64194989R, (01)00889024207790(17)281130(10)64202115, (01)00889024207790(17)281130(10)64226003, (01)00889024207806(17)281130(10)64226007, (01)00889024207806(17)281130(10)64226008, (01)00889024207820(17)281130(10)64220651, (01)00889024207790(17)281130(10)64226001, (01)00889024207806(17)281130(10)64226006, (01)00889024207806(17)281130(10)64202119, (01)00889024207790(17)281130(10)64226002, (01)00889024207806(17)281130(10)64202118, (01)00889024207790(17)281231(10)64254389, (01)00889024207806(17)281231(10)64254390, (01)00889024207806(17)281231(10)64226077, (01)00889024207806(17)281231(10)64226078, (01)00889024207806(17)281231(10)64226009, (01)00889024207790(17)281231(10)64226076, (01)00889024207790(17)281231(10)64254387, (01)00889024207790(17)281231(10)64254388, (01)00889024207806(17)281231(10)64261032, (01)00889024207806(17)281231(10)64226129, (01)00889024207820(17)281231(10)64242244, (01)00889024207790(17)281231(10)64254386, (01)00889024207790(17)281231(10)64226076R, (01)00889024207790(17)281231(10)64254386R, (01)00889024207790(17)281231(10)64254387R, (01)00889024207790(17)290131(10)64261030, (01)00889024207790(17)290131(10)64275592, (01)00889024207806(17)290131(10)64275434, (01)00889024207806(17)290131(10)64261033, (01)00889024207806(17)290131(10)64261035, (01)00889024207790(17)290131(10)64261031, (01)00889024207806(17)290131(10)64275593, (01)00889024207790(17)290131(10)64261029, (01)00889024207790(17)290131(10)64275538, (01)00889024207806(17)290131(10)64275540, (01)00889024207820(17)290131(10)64261037, (01)00889024207813(17)290131(10)64261036, (01)00889024207790(17)290131(10)64275537, (01)00889024207806(17)290131(10)64275435, (01)00889024207790(17)290131(10)64275433, (01)00889024207806(17)290131(10)64275541, (01)00889024207813(17)290131(10)64261036R, (01)00889024207790(17)290131(10)64261031R, (01)00889024207806(17)290131(10)64275593R, (01)00889024207790(17)290228(10)64294621, (01)00889024207820(17)290228(10)64306997, (01)00889024207790(17)290228(10)64302414, (01)00889024207790(17)290228(10)64306991, (01)00889024207806(17)290228(10)64302416, (01)00889024207806(17)290228(10)64306994, (01)00889024207806(17)290228(10)64302415, (01)00889024207790(17)290228(10)64275591, (01)00889024207790(17)290228(10)64275432, (01)00889024207806(17)290228(10)64306994R, (01)00889024207806(17)290228(10)64302415R, (01)00889024207790(17)290228(10)64302414R, (01)00889024207806(17)290331(10)64346982, (01)00889024207806(17)290331(10)64334419, (01)00889024207806(17)290331(10)64306995, (01)00889024207806(17)290331(10)64261034, (01)00889024207790(17)290331(10)64306992, (01)00889024207790(17)290331(10)64306993, (01)00889024207813(17)290331(10)64306996, (01)00889024207806(17)290331(10)64334418, (01)00889024207806(17)290331(10)64334420, (01)00889024207790(17)290331(10)64306992R, (01)00889024207790(17)290331(10)64306993R, (01)00889024207813(17)290331(10)64306996R, (01)00889024207790(17)290430(10)64346981, (01)00889024207790(17)290430(10)64367312, (01)00889024207806(17)290430(10)64367404, (01)00889024207806(17)290430(10)64346996, (01)00889024207806(17)290430(10)64367314, (01)00889024207790(17)290430(10)64334415, (01)00889024207. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Out of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide distribution - US Nationwide including the states of AK, AL, CA, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the countries of Australia, Canada, Japan, Netherlands.
- Lot
- Item No. 00595201501, 00595201502, 00595201505, 00595201506; UDI:(01)00889024207806(17)280731(10)64115713, (01)00889024207806(17)280930(10)64175075, (01)00889024207806(17)280930(10)64175076, (01)00889024207790(17)280930(10)64175073, (01)00889024207806(17)280930(10)64138317, (01)00889024207790(17)281031(10)64175074, (01)00889024207790(17)281031(10)64194991, (01)00889024207790(17)281031(10)64202116, (01)00889024207806(17)281031(10)64178998, (01)00889024207813(17)281031(10)64179054, (01)00889024207790(17)281031(10)64202117, (01)00889024207813(17)281031(10)64179043, (01)00889024207813(17)281031(10)64194989, (01)00889024207790(17)281031(10)64178996, (01)00889024207790(17)281031(10)64178997, (01)00889024207790(17)281031(10)64179023, (01)00889024207820(17)281031(10)64178999, (01)00889024207813(17)281031(10)64179043R, (01)00889024207813(17)281031(10)64194989R, (01)00889024207790(17)281130(10)64202115, (01)00889024207790(17)281130(10)64226003, (01)00889024207806(17)281130(10)64226007, (01)00889024207806(17)281130(10)64226008, (01)00889024207820(17)281130(10)64220651, (01)00889024207790(17)281130(10)64226001, (01)00889024207806(17)281130(10)64226006, (01)00889024207806(17)281130(10)64202119, (01)00889024207790(17)281130(10)64226002, (01)00889024207806(17)281130(10)64202118, (01)00889024207790(17)281231(10)64254389, (01)00889024207806(17)281231(10)64254390, (01)00889024207806(17)281231(10)64226077, (01)00889024207806(17)281231(10)64226078, (01)00889024207806(17)281231(10)64226009, (01)00889024207790(17)281231(10)64226076, (01)00889024207790(17)281231(10)64254387, (01)00889024207790(17)281231(10)64254388, (01)00889024207806(17)281231(10)64261032, (01)00889024207806(17)281231(10)64226129, (01)00889024207820(17)281231(10)64242244, (01)00889024207790(17)281231(10)64254386, (01)00889024207790(17)281231(10)64226076R, (01)00889024207790(17)281231(10)64254386R, (01)00889024207790(17)281231(10)64254387R, (01)00889024207790(17)290131(10)64261030, (01)00889024207790(17)290131(10)64275592, (01)00889024207806(17)290131(10)64275434, (01)00889024207806(17)290131(10)64261033, (01)00889024207806(17)290131(10)64261035, (01)00889024207790(17)290131(10)64261031, (01)00889024207806(17)290131(10)64275593, (01)00889024207790(17)290131(10)64261029, (01)00889024207790(17)290131(10)64275538, (01)00889024207806(17)290131(10)64275540, (01)00889024207820(17)290131(10)64261037, (01)00889024207813(17)290131(10)64261036, (01)00889024207790(17)290131(10)64275537, (01)00889024207806(17)290131(10)64275435, (01)00889024207790(17)290131(10)64275433, (01)00889024207806(17)290131(10)64275541, (01)00889024207813(17)290131(10)64261036R, (01)00889024207790(17)290131(10)64261031R, (01)00889024207806(17)290131(10)64275593R, (01)00889024207790(17)290228(10)64294621, (01)00889024207820(17)290228(10)64306997, (01)00889024207790(17)290228(10)64302414, (01)00889024207790(17)290228(10)64306991, (01)00889024207806(17)290228(10)64302416, (01)00889024207806(17)290228(10)64306994, (01)00889024207806(17)290228(10)64302415, (01)00889024207790(17)290228(10)64275591, (01)00889024207790(17)290228(10)64275432, (01)00889024207806(17)290228(10)64306994R, (01)00889024207806(17)290228(10)64302415R, (01)00889024207790(17)290228(10)64302414R, (01)00889024207806(17)290331(10)64346982, (01)00889024207806(17)290331(10)64334419, (01)00889024207806(17)290331(10)64306995, (01)00889024207806(17)290331(10)64261034, (01)00889024207790(17)290331(10)64306992, (01)00889024207790(17)290331(10)64306993, (01)00889024207813(17)290331(10)64306996, (01)00889024207806(17)290331(10)64334418, (01)00889024207806(17)290331(10)64334420, (01)00889024207790(17)290331(10)64306992R, (01)00889024207790(17)290331(10)64306993R, (01)00889024207813(17)290331(10)64306996R, (01)00889024207790(17)290430(10)64346981, (01)00889024207790(17)290430(10)64367312, (01)00889024207806(17)290430(10)64367404, (01)00889024207806(17)290430(10)64346996, (01)00889024207806(17)290430(10)64367314, (01)00889024207790(17)290430(10)64334415, (01)00889024207
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAVanguard Knee System AS Tibial Bearing / DCM ArCom, 12 MM X 75 MM; Model Number 189082. The product is placedMay 1, 2024
- FDARoute 92 Medical Full Length 070 Access System, REF 7002, Rxonly, SterileEO. Used for Introduction ofMay 1, 2024
Questions about this recall
Was NexGen CR-Flex Femoral Component, Size E and Size E Minus (E-). Component of the NexGen CR-Flex Porous Femur System. recalled?
Yes. FDA published recall Z-1608-2024 on May 1, 2024. Reported hazard: Out of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.
What should I do if I have this product?
Voluntary: Firm initiated · Letter