t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology
Published May 15, 2024 · Recall initiated: March 5, 2024
Summary
FDA records medical device Class I recall Z-1609-2024, published May 15, 2024 and initiated March 5, 2024. Recalling firm: Tandem Diabetes Care, Inc.; Status: Ongoing; Reported quantity: 85,863 applications; Lot: version 2.7 or later /UDI: 00850006613274. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Tandem Diabetes Care, Inc.
- Distributor
- US nationwide distribution including Puerto Rico.
- Lot
- version 2.7 or later /UDI: 00850006613274
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700,November 11, 2026
- FDABrand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/CatalogOctober 7, 2026
- FDAt:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072,October 29, 2025
- FDABelgian Yummies Homemade Ice Cream Horchata Gelato. Packaged in 4.75 Liter gray pan.May 15, 2024
- FDAVyaire Medical GmbH, Vyaire Twin Tube sample line 240 cm, Model Number V-707327; The TwinTube sample lineMay 15, 2024
- FDAPeanut Butter Ice Cream Cookie Sandwich. Sandwiches packaged in a gold or clear bags. 50 sandwiches per caseMay 15, 2024
Questions about this recall
Was t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology recalled?
Yes. FDA published recall Z-1609-2024 on May 15, 2024. Reported hazard: During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.
What should I do if I have this product?
Voluntary: Firm initiated · Letter