Philips V60 Ventilators with Power Management PCBA part number 1055906
Published April 8, 2020 · Recall initiated: March 5, 2020
Summary
FDA records medical device Class II recall Z-1622-2020, published April 8, 2020 and initiated March 5, 2020. Recalling firm: Respironics California, LLC; Status: Ongoing; Reported quantity: 23581; Lot: Serial Numbers: 100002908 to 100027093; 100027096 to 100085389; 100085399 to 100109746; 100109868 to 201003390; 201003392 to 201010952. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Respironics California, LLC
- Distributor
- U.S.: CO, PA, AZ, OH, MT, AR, FL, NC, KS, IL, CA, HI, MD, SC, WY, OR, NY, NE, OK, MN, UT, ND, LA, IN, AL, AK, WV, GA, WI, MI, VA, NJ, TX, WA, SD, KY, TN, MS, MO, DE, MA, VT, CT, IA, NV, ID, ME, NM, DC, NH, VI, RI, PR. O.U.S.: Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Gibraltar, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jersey, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Malta, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R®union, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam
- Lot
- Serial Numbers: 100002908 to 100027093; 100027096 to 100085389; 100085399 to 100109746; 100109868 to 201003390; 201003392 to 201010952
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Respironics V680 Ventilator (All Models, including: P/N 1103044 (850011)), *Distributed outside theApril 27, 2022
- FDAPhilips Respironics V60 Plus Ventilator (All Models, including: 1138747, 1137276)April 27, 2022
- FDAPhilips Respironics V60 Plus Ventilator Part Number 1138747March 9, 2022
- FDAPhilips Respironics V60 Ventilator Part Number 1053614 - Product Usage: intended to be used only with variousFebruary 3, 2021
- FDAKarl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series,April 8, 2020
- FDAGentell Hydrogel Ag 4x4, model no. GEN-11420 - Product Usage: Under the care of a healthcare professional,April 8, 2020
Questions about this recall
Was Philips V60 Ventilators with Power Management PCBA part number 1055906 recalled?
Yes. FDA published recall Z-1622-2020 on April 8, 2020. Reported hazard: Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
What should I do if I have this product?
Voluntary: Firm initiated · Letter