QUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20, 20 Tests

Published May 1, 2024 · Recall initiated: March 5, 2024

Summary

FDA records medical device Class II recall Z-1637-2024, published May 1, 2024 and initiated March 5, 2024. Recalling firm: Quidel Corporation; Status: Ongoing; Reported quantity: 31,140 units; Lot: UDI: 30014613336747/Lot # 162620 and 162748. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Quidel Corporation
Distributor
US: AK AL AR AZ CA CO CT DC FL GA HI IA ID IL IN KS KY LA MA MD MI MN MO MT NC ND NE NH NJ NV NY OH OK OR PA SC SD TN TX UT VA WA WI WV OUS: United Arab Emirates, Belgium, Canada, Chile, China, Germany, Spain, United Kingdom, Ireland, Italy, South Korea, Lithuania, Qatar
Lot
UDI: 30014613336747/Lot # 162620 and 162748
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was QUIDEL InflammaDry MMP-9 Test, REF: PRS-ID-20, 20 Tests recalled?

Yes. FDA published recall Z-1637-2024 on May 1, 2024. Reported hazard: There are two windows on the test device, the control line window and the test results window. The line for the test results window were misaligned. In instances of positive test results, the positivite test line may not be visible/ hidden under the cassette of the test device thus potentially leading to a false negative result.

What should I do if I have this product?

Voluntary: Firm initiated · Letter