Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +0 Neck, Model Number 802802202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented

Published April 30, 2025 · Recall initiated: March 24, 2025

Summary

FDA records medical device Class II recall Z-1648-2025, published April 30, 2025 and initiated March 24, 2025. Recalling firm: Zimmer, Inc.; Status: Ongoing; Reported quantity: 158 units; Lot: GTIN 00887868376351, All lots distributed with IFU 87-6204-051-99 Rev D or prior. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.
Lot
GTIN 00887868376351, All lots distributed with IFU 87-6204-051-99 Rev D or prior
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +0 Neck, Model Number 802802202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented recalled?

Yes. FDA published recall Z-1648-2025 on April 30, 2025. Reported hazard: The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit