FDAFDA · Food and drugs·Recall number Z-1651-2014

Gluma Desensitizer Power Gel Product Usage: Usage: This is a class 2 medical device. For reduction or elimination of pain in exposed cervical areas that do not require restoration. It is also used for alleviation or reduction of dentinal sensitivity after preparation of teeth to receive fixed prosthesis or restorations.

Published May 28, 2014 · Recall initiated: April 23, 2014

Summary

FDA records medical device Class II recall Z-1651-2014, published May 28, 2014 and initiated April 23, 2014. Recalling firm: Heraeus Kulzer, LLC.; Status: Terminated; Reported quantity: 5,112 syringes or 1,278 kits (1 Kit=4 syringes); Lot: Model Number(s): 66043451 Expiry: 2015-09 Syringe Lot number 010103, which are contained within Kit lot numbers 42.43.44.45.46.47.48.49,50, and 51. Note; lots 49 and 50 were captured in house, 0 was distributed. Lot 51 was made by the manufacturer, however, never received by us or distributed. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm discovered that the Gluma Desensitizer Power Gel can be inadvertently extruded from the application syringe all at once when it is supposed to extrude on the desired area a small amount at a time. If extruded all at once, product could reach the dental patients mucosa, face or eyes or the dental professional could be exposed. This would cause irritation, burning or blistering.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Heraeus Kulzer, LLC.
Distributor
US Nationwide Distribution in the states of GA, TX, IL, CA, WA, WI, HI, MO, FL, NJ, TN, PA, NY, NC, OH, IN, CT, KY, AZ, NV, IA,
Lot
Model Number(s): 66043451 Expiry: 2015-09 Syringe Lot number 010103, which are contained within Kit lot numbers 42.43.44.45.46.47.48.49,50, and 51. Note; lots 49 and 50 were captured in house, 0 was distributed. Lot 51 was made by the manufacturer, however, never received by us or distributed.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Gluma Desensitizer Power Gel Product Usage: Usage: This is a class 2 medical device. For reduction or elimination of pain in exposed cervical areas that do not require restoration. It is also used for alleviation or reduction of dentinal sensitivity after preparation of teeth to receive fixed prosthesis or restorations. recalled?

Yes. FDA published recall Z-1651-2014 on May 28, 2014. Reported hazard: The firm discovered that the Gluma Desensitizer Power Gel can be inadvertently extruded from the application syringe all at once when it is supposed to extrude on the desired area a small amount at a time. If extruded all at once, product could reach the dental patients mucosa, face or eyes or the dental professional could be exposed. This would cause irritation, burning or blistering.

What should I do if I have this product?

Voluntary: Firm initiated · Letter