FDAFDA · Food and drugs·Recall number Z-1659-2016
Dimension Assays: Direct HDL Cholesterol (AHDL)
Published May 18, 2016 · Recall initiated: March 17, 2016
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific
What to do
Voluntary: Firm initiated
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands
- Lot
- Device Listing: D 044702 - Direct HDL Cholesterol (AHDL), Catalog # DF48B
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States