FDAFDA · Food and drugs·Recall number Z-1661-2015
Direct Water feed Kit, Catalog#10373222
Published June 3, 2015 · Recall initiated: March 5, 2015
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the cap from closing or fastening securely to the bottle.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Medical Solutions Diagnostics
- Distributor
- Worldwide Distribution.
- Lot
- Direct Water Feed Kit - Product Code/Catalog #10373222
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States