ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;
Published April 8, 2026 · Recall initiated: March 3, 2026
Summary
FDA records medical device Class II recall Z-1664-2026, published April 8, 2026 and initiated March 3, 2026. Recalling firm: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Status: Ongoing; Reported quantity: 461 units (198 US, 263 OUS); Lot: (1) System Code: 722008; UDI: N/A; System Serial Number: 360, 361, 445, 296, 487, 390, 306, 178, 420, 267, 245, 2470, 491, 113, 386, 69, 413, 365, 454, 304, 442, 125, 441, 320, 325, 112, 254, 456, 52, 34, 239, 407, 187, 437, 82, 102, 288, 349, 106, 105, 481, 71, 424, 1236, 197, 264, 18, 19, 192, 124, 483, 244, 224, 470, 279, 141, 459, 357, 287, 116, 521, 369, 212, 23, 415, 100, 294, 270, 372, 472, 18270, 154, 231, 108, 1178, 246, 115, 460, 319, 465, 142, 419, 312, 37, 59, 455, 217, 350, 406, 160, 167, 435, 447, 410, 196, 458, 480, 476, 370, 169, 121, 259, 266, 186, 290, 253, 359, 391, 283, 205, 425, 346, 376, 150, 394, 123, 513, 337, 93, 268, 200, 87, 110, 46, 542, 334, 122, 8, 468, 250, 482, 63, 408, 305, 213, 547, 546, 475, 356, 404, 398, 31, 553; (2) System Code: 722013; UDI: 00884838059061; System Serial Number: 293, 424, 100, 174, 86, 83, 46, 170, 218, 441, 249, 639, 290, 153, 410, 1, 289, 518, 519, 626, 126, 141, 496, 379, 115, 42, 601, 214, 637, 360, 389, 303, 268, 264, 631, 254, 255, 515, 129, 15, 603, 278, 347, 426, 505, 109, 462, 149, 124, 216, 246, 84, 479, 69, 105, 638, 375, 531, 404, 575, 147, 475, 262, 125, 316, 497, 357, 495, 573, 342, 564, 164, 178, 556, 253, 297, 511, 492, 43, 550, 96, 179, 228, 194, 151, 106, 330, 461, 304, 40, 85, 72, 241, 422, 267, 150, 428, 634, 327, 442, 175, 217, 313, 346, 308, 562, 309, 152, 34, 561, 35, 205, 61, 435, 266, 567, 168, 607, 283, 52, 171, 66, 36, 585, 271, 117, 443, 107, 138, 223, 343, 165, 453, 370, 367, 108, 352, 291, 259, 455, 368, 315, 325, 532, 598, 440, 172, 193, 301, 415, 540, 62, 299, 183, 184, 185, 186, 187, 188, 189, 190, 413, 437, 560, 300, 295, 448, 81, 444, 400, 110, 111, 624, 65, 633, 630, 361, 549, 417, 449, 506, 177, 629, 476, 527, 276, 247, 156, 499, 50, 302, 618, 608, 133, 70, 457, 547, 312, 366, 390, 544, 236, 157, 243, 144, 238, 525, 53, 570, 161, 349, 359, 454, 613, 623, 222, 215, 335, 445, 538, 416, 628, 619, 47, 206, 620, 384, 526, 332, 358, 470, 378, 296, 227, 31, 58, 589, 616, 71, 447, 530, 232, 324, 311, 176, 181, 554, 134, 163, 104, 403, 478, 459, 17, 528, 318, 28, 237, 559, 372, 256, 374, 487, 405, 380, 434, 130, 473, 551, 557, 294, 412, 128, 388, 207, 258, 429, 95, 411, 353, 542, 533, 622, 182, 574, 139, 137, 250, 563, 502, 450, 436, 406, 354, 510, 395, 458, 261, 348, 503, 595, 586, 535, 569, 425, 334, 344, 355, 465, 345, 522, 64, 208, 408, 555, 394, 57, 493. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distributor
- Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C®te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, FrenchPolynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Laos, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R®union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkiye, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen;
- Lot
- (1) System Code: 722008; UDI: N/A; System Serial Number: 360, 361, 445, 296, 487, 390, 306, 178, 420, 267, 245, 2470, 491, 113, 386, 69, 413, 365, 454, 304, 442, 125, 441, 320, 325, 112, 254, 456, 52, 34, 239, 407, 187, 437, 82, 102, 288, 349, 106, 105, 481, 71, 424, 1236, 197, 264, 18, 19, 192, 124, 483, 244, 224, 470, 279, 141, 459, 357, 287, 116, 521, 369, 212, 23, 415, 100, 294, 270, 372, 472, 18270, 154, 231, 108, 1178, 246, 115, 460, 319, 465, 142, 419, 312, 37, 59, 455, 217, 350, 406, 160, 167, 435, 447, 410, 196, 458, 480, 476, 370, 169, 121, 259, 266, 186, 290, 253, 359, 391, 283, 205, 425, 346, 376, 150, 394, 123, 513, 337, 93, 268, 200, 87, 110, 46, 542, 334, 122, 8, 468, 250, 482, 63, 408, 305, 213, 547, 546, 475, 356, 404, 398, 31, 553; (2) System Code: 722013; UDI: 00884838059061; System Serial Number: 293, 424, 100, 174, 86, 83, 46, 170, 218, 441, 249, 639, 290, 153, 410, 1, 289, 518, 519, 626, 126, 141, 496, 379, 115, 42, 601, 214, 637, 360, 389, 303, 268, 264, 631, 254, 255, 515, 129, 15, 603, 278, 347, 426, 505, 109, 462, 149, 124, 216, 246, 84, 479, 69, 105, 638, 375, 531, 404, 575, 147, 475, 262, 125, 316, 497, 357, 495, 573, 342, 564, 164, 178, 556, 253, 297, 511, 492, 43, 550, 96, 179, 228, 194, 151, 106, 330, 461, 304, 40, 85, 72, 241, 422, 267, 150, 428, 634, 327, 442, 175, 217, 313, 346, 308, 562, 309, 152, 34, 561, 35, 205, 61, 435, 266, 567, 168, 607, 283, 52, 171, 66, 36, 585, 271, 117, 443, 107, 138, 223, 343, 165, 453, 370, 367, 108, 352, 291, 259, 455, 368, 315, 325, 532, 598, 440, 172, 193, 301, 415, 540, 62, 299, 183, 184, 185, 186, 187, 188, 189, 190, 413, 437, 560, 300, 295, 448, 81, 444, 400, 110, 111, 624, 65, 633, 630, 361, 549, 417, 449, 506, 177, 629, 476, 527, 276, 247, 156, 499, 50, 302, 618, 608, 133, 70, 457, 547, 312, 366, 390, 544, 236, 157, 243, 144, 238, 525, 53, 570, 161, 349, 359, 454, 613, 623, 222, 215, 335, 445, 538, 416, 628, 619, 47, 206, 620, 384, 526, 332, 358, 470, 378, 296, 227, 31, 58, 589, 616, 71, 447, 530, 232, 324, 311, 176, 181, 554, 134, 163, 104, 403, 478, 459, 17, 528, 318, 28, 237, 559, 372, 256, 374, 487, 405, 380, 434, 130, 473, 551, 557, 294, 412, 128, 388, 207, 258, 429, 95, 411, 353, 542, 533, 622, 182, 574, 139, 137, 250, 563, 502, 450, 436, 406, 354, 510, 395, 458, 261, 348, 503, 595, 586, 535, 569, 425, 334, 344, 355, 465, 345, 522, 64, 208, 408, 555, 394, 57, 493;
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDAPhilips Azurion 5 M12. Model Number: 722227.November 4, 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.November 4, 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.November 4, 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.November 4, 2026
- FDAImu-Tek Colostrum-5 Powder; 30% IgG,1,000 mg 2x/day, orally with 8 ounces of water, support a healthy immuneApril 8, 2026
- FDAArtoura Breast Tissue Expanders Reference Numbers: TEXP110RH TEXP120RH TEXP120RUH TEXP130RH TEXP135RHApril 8, 2026
Questions about this recall
Was ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013; recalled?
Yes. FDA published recall Z-1664-2026 on April 8, 2026. Reported hazard: Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
What should I do if I have this product?
Voluntary: Firm initiated · Letter