Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO CLS3.5 (5PK), REF H74908526574A2; cardiac catheter
Published May 22, 2024 · Recall initiated: March 21, 2024
Summary
FDA records medical device Class I recall Z-1678-2024, published May 22, 2024 and initiated March 21, 2024. Recalling firm: Boston Scientific Corporation; Status: Ongoing; Reported quantity: 615 catheters; Lot: UDI/DI 08714729812203, Batch Numbers: 60447147, 60467099, 60492626. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- Worldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions.
- Lot
- UDI/DI 08714729812203, Batch Numbers: 60447147, 60467099, 60492626
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDANaviFlex LW Pusher; Advx Panc. Stent 4/5F; Material Number (UPN): M00535030;October 21, 2026
- FDAImager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5;September 30, 2026
- FDABoston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVISeptember 23, 2026
- FDALUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS US, Model NumberSeptember 9, 2026
- FDAExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO CLS3 (5PK), REF H74908526573A2; cardiac catheterMay 22, 2024
- FDAExpo 5F Selective Angiographic Catheters, MODEL - 5F EXPO CLS4 (5PK), REF H74908526575A2; cardiac catheterMay 22, 2024
Questions about this recall
Was Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO CLS3.5 (5PK), REF H74908526574A2; cardiac catheter recalled?
Yes. FDA published recall Z-1678-2024 on May 22, 2024. Reported hazard: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.
What should I do if I have this product?
Voluntary: Firm initiated · Letter