FDAFDA · Food and drugs·Recall number Z-1679-2014

115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.

Published June 11, 2014 · Recall initiated: January 8, 2014

Summary

FDA records medical device Class II recall Z-1679-2014, published June 11, 2014 and initiated January 8, 2014. Recalling firm: Cincinnati Sub-Zero Products Inc; Status: Terminated; Reported quantity: 7 units; Lot: Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cincinnati Sub-Zero Products Inc
Distributor
The product was distributed to one customer in MI, and internationally (Taiwan)
Lot
Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was 115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165. recalled?

Yes. FDA published recall Z-1679-2014 on June 11, 2014. Reported hazard: The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but

What should I do if I have this product?

Voluntary: Firm initiated · Telephone