115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165.
Published June 11, 2014 · Recall initiated: January 8, 2014
Summary
FDA records medical device Class II recall Z-1679-2014, published June 11, 2014 and initiated January 8, 2014. Recalling firm: Cincinnati Sub-Zero Products Inc; Status: Terminated; Reported quantity: 7 units; Lot: Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but
What to do
Voluntary: Firm initiated · Telephone
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cincinnati Sub-Zero Products Inc
- Distributor
- The product was distributed to one customer in MI, and internationally (Taiwan)
- Lot
- Model Number 222S; Serial #s: 133-2-00697, 133-2-00698; 133-2-00699; 133-2-00700; 133-2-00701; 133-2-00702; and 133-2-00703:.Part # 86165
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHemotherm CE, model 400CE, Heater-Cooling DeviceApril 26, 2017
- FDAIntelliSpace Portal (ISP), (ISP DX/HX/EX, Model number 881001; ISP IX, Model number 881030; ISP LX SPECT,June 11, 2014
- FDAWound Drain - Label reads in part "Elutia ***Hubless Coated Silicone Surgical Wound Drain***" Two differentJune 11, 2014
- FDAenGen (TM) Laboratory Automation System configured with Data Innovations, LLC Instrument Manager (TM) (IM)June 11, 2014
- FDABlanketrol III Data Export Software Version 1.10 The Blanketrol III Data Export Software is intended for useDecember 11, 2013
- FDABlanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OHSeptember 25, 2013
Questions about this recall
Was 115 V Blanketrol II, Model 222S Hyper-Hypothermia System . Used to lower or to raise a patients temperature and /or maintain a desired patient temperature through conductive heat transfer. Part number 86165. recalled?
Yes. FDA published recall Z-1679-2014 on June 11, 2014. Reported hazard: The incorrect voltage component was placed into the device during manufacturing. This may make the device unable to maintain patient temperature during cooling mode. The device will continue to indicate that the unit is cooling because the compressor will continue to run, however water in the reservoir will begin to return to ambient temperature. The actual water temperature will be displayed but
What should I do if I have this product?
Voluntary: Firm initiated · Telephone