IMMULITE ¿ /IMMULITE 1000 ¿ Gastrin
Published May 16, 2018 · Recall initiated: January 2, 2018
Summary
FDA records medical device Class II recall Z-1680-2018, published May 16, 2018 and initiated January 2, 2018. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: N/A; Lot: Gastrin, Gastrin (IMMULITE/IMMULITE 1000, LKGA1), (IMMULITE 2000/IMMULITE 2000 XPi, LKGA2) Lot codes: Gastrin Lot # 0345 0346 0347. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Nationwide and Canada, Mexico, and Thailand
- Lot
- Gastrin, Gastrin (IMMULITE/IMMULITE 1000, LKGA1), (IMMULITE 2000/IMMULITE 2000 XPi, LKGA2) Lot codes: Gastrin Lot # 0345 0346 0347
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAIMMULITE ¿ /IMMULITE ¿ 1000 BR-MA (CA15-3)May 16, 2018
- FDAIMMULITE ¿ 2000/IMMULITE ¿ 2000 XPi GastrinMay 16, 2018
Questions about this recall
Was IMMULITE ¿ /IMMULITE 1000 ¿ Gastrin recalled?
Yes. FDA published recall Z-1680-2018 on May 16, 2018. Reported hazard: Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
What should I do if I have this product?
Voluntary: Firm initiated · Letter