FDAFDA · Food and drugs·Recall number Z-1680-2024

Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AR1 (5PK), REF H74908526842; cardiac catheter

Published May 22, 2024 · Recall initiated: March 21, 2024

Summary

FDA records medical device Class I recall Z-1680-2024, published May 22, 2024 and initiated March 21, 2024. Recalling firm: Boston Scientific Corporation; Status: Ongoing; Reported quantity: 11109 catheters; Lot: UDI/DI 08714729187929, Batch Numbers: 60425143, 60448984, 60474111, 60482880, 60525062, 60427595, 60448985, 60477260, 60502674, 60530043, 60429297, 60463971, 60479435, 60511249, 60536205. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Boston Scientific Corporation
Distributor
Worldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions.
Lot
UDI/DI 08714729187929, Batch Numbers: 60425143, 60448984, 60474111, 60482880, 60525062, 60427595, 60448985, 60477260, 60502674, 60530043, 60429297, 60463971, 60479435, 60511249, 60536205
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AR1 (5PK), REF H74908526842; cardiac catheter recalled?

Yes. FDA published recall Z-1680-2024 on May 22, 2024. Reported hazard: An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

What should I do if I have this product?

Voluntary: Firm initiated · Letter