Cyto-Chex BCT - blood specimen collection device - Immunophenotyping Preservative. IVD. Sterile. 2.0ml only. Packaged 6-tube pack REF 213559 and 100-tube box REF 218980. Streck.
Published April 5, 2017 · Recall initiated: February 23, 2017
Summary
FDA records medical device Class II recall Z-1684-2017, published April 5, 2017 and initiated February 23, 2017. Recalling firm: Streck; Status: Ongoing; Reported quantity: 128,692 tubes; Lot: UDI Number 10844509000836 and 20844509000833. Lot(s) Expiration Date 50910309 04-01-2017 60760309 03-16-2018 61460309 05-25-2018. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Streck
- Distributor
- Worldwide Distribution - USA (nationwide) and to the countries of : Belgium, Canada, Germany, Israel, Japan, the Republic of Korea, the Netherlands, and the United Kingdom. There was no military or government distribution.
- Lot
- UDI Number 10844509000836 and 20844509000833. Lot(s) Expiration Date 50910309 04-01-2017 60760309 03-16-2018 61460309 05-25-2018
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPlum 360 Infusion Pump, Software Version 15.02. The infusion pump is capable of delivering fluids for aApril 5, 2017
- FDALegionella IFA Substrate Slide (IF0951) intended for qualitatively detecting and semi-quantitating human serumApril 5, 2017
- FDAMerge Eye Care Systems, consisting of Merge Eye Station, Merge Eye Care PACS, and Merge Eye Station ImportApril 5, 2017
- FDASyngo.x, Picture archiving and communication system Syngo via is a software solution intended to be used forApril 5, 2017
- FDAJResultNet drivers using ASTM coding language. Calculator/data processing module, for clinical use -April 5, 2017
- FDAUOC Femoral Driver, Product number PE 93045101, orthopedic manual surgical instrument/accessory for use duringApril 5, 2017
Questions about this recall
Was Cyto-Chex BCT - blood specimen collection device - Immunophenotyping Preservative. IVD. Sterile. 2.0ml only. Packaged 6-tube pack REF 213559 and 100-tube box REF 218980. Streck. recalled?
Yes. FDA published recall Z-1684-2017 on April 5, 2017. Reported hazard: Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit