FDAFDA · Food and drugs·Recall number Z-1690-2024
Philips Patient Information Center (PIC) iX Uninterruptable Power Supply (UPS): 1. APC (Schneider Electric IT Corporation) 120V Uninterruptable Power Supply Model: SCL500RM1U 2. APC (Schneider Electric IT Corporation) 230V Uninterruptable Power Supply Model: SCL500RMI1U
Published May 1, 2024 · Recall initiated: March 22, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Possible failure of Uninterruptable Power Supply (UPS) devices supporting the Patient Information Center (PIC) iX system, potential for the associated PIC iX system to shut down due to the lack of power. may lead to the delay in the detection of a change in a patient condition
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, Estonia, Fiji, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Oman, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, United Kingdom, Utd. Ara Emir.
- Lot
- 1. SCL500RM1U (only serial number date codes prior to 2237*) 2. SCL500RMI1U (only serial number date codes prior to 2241*) Example unit serial number is 5S"2104"08943, meaning it was manufactured during week 04 of the year 2021. The serial number format is 5SYYWW123456.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States