FDAFDA · Food and drugs·Recall number Z-1698-2021

T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to insert screws with T25 driving features, which includes screws with diameters of 5.5mm, 6.5mm, and 7.3mm Product Code: GWK 300

Published June 9, 2021 · Recall initiated: April 15, 2021

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Cannulated Screwdrivers may break at the driver tip along the shaft, near the handle extending surgery time and additional x-ray to identify the fragments from the breakages

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Flower Orthopedics Corporation
Distributor
US Nationwide distribution in the states of AZ, CO, FL, GA, IL, IN, PA, NC, VA.
Lot
Lot Numbers: 2003320898, 1810220041, 1802220008, 1802220005, 1711220035 UDI: 00840118110853
Risk classification
Class II
Area (FDA)
device
Country
United States