FDAFDA · Food and drugs·Recall number Z-1705-2013
REF 154214 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size C3 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.
Published July 17, 2013 · Recall initiated: June 7, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands.
- Lot
- M321570, M894590,M619770, M480930, M656950 M444780, M969100
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States