REF 154216 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size C5 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.
Published July 17, 2013 · Recall initiated: June 7, 2013
Summary
FDA records medical device Class II recall Z-1707-2013, published July 17, 2013 and initiated June 7, 2013. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 23; Lot: M539400, M619790, M987360 , M444800, M969130, M151150, M987360. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet, Inc.
- Distributor
- Worldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands.
- Lot
- M539400, M619790, M987360 , M444800, M969130, M151150, M987360
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475640 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475625 Product Description: Intramedullary BoneAugust 12, 2026
- FDAIntramedullary Bone Saw Blade Assembly Model/Catalog Number: 475630 Product Description: Intramedullary BoneAugust 12, 2026
- FDAREF 154215 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size C4 ARCOM UHMWPE / CO-CR-MO Alloy OnlyJuly 17, 2013
- FDAREF 154217 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size C6 ARCOM UHMWPE / CO-CR-MO Alloy OnlyJuly 17, 2013
Questions about this recall
Was REF 154216 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size C5 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement. recalled?
Yes. FDA published recall Z-1707-2013 on July 17, 2013. Reported hazard: Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.
What should I do if I have this product?
Voluntary: Firm initiated · Letter