AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.
Published June 11, 2014 · Recall initiated: April 11, 2014
Summary
FDA records medical device Class II recall Z-1711-2014, published June 11, 2014 and initiated April 11, 2014. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 37; Lot: Model number: 10094200, serial numbers:4332 4291 4323 4430 3061 4331 4238 4297 4201 4024 3098 4252 4372 4354 4077 3050 3108 4327 4342 4251 3053 3114 4156 3145 3096 4043 3181 4208 3090 4143 4051 4199 4401 4405 4151 4373 4276. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Class II Recall - US Distribution including the states of IL,VA, TN, PA, TX, MN, GA, MI, IN, OH, FL, CA, OR, NY, ME, MO, OK, CO and NJ.
- Lot
- Model number: 10094200, serial numbers:4332 4291 4323 4430 3061 4331 4238 4297 4201 4024 3098 4252 4372 4354 4077 3050 3108 4327 4342 4251 3053 3114 4156 3145 3096 4043 3181 4208 3090 4143 4051 4199 4401 4405 4151 4373 4276
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDALiquid Mouse Monoclonal Antibody CD15 (NCL-L-CD15), a liquid tissue culture supernatant containing 15 mMJune 11, 2014
- FDAPATH(R) THREAD CUP ADAPTOR, REF 2007-0013, Wright Medical Technology, Inc. Orthopaedic surgical instrument.June 11, 2014
Questions about this recall
Was AXIOM Luminos dRFsystem The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. recalled?
Yes. FDA published recall Z-1711-2014 on June 11, 2014. Reported hazard: It was determined that under rare environmental conditions (i.e., extremely dry air in combination with a highly isolated floor) electrostatic discharges (ESD) >8 kV may trigger an unintended movement of the AXIOM Luminos dRFsystem which may lead to an emergency situation and pose danger to the patients, operating personnel or to the unit.
What should I do if I have this product?
Voluntary: Firm initiated · Letter