ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c
Published April 12, 2017 · Recall initiated: February 23, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution - United States Nationwide: Denmark Sweden Norway Great Britain Belgium Netherlands Spain Israel Switzerland France Portugal Greece Germany Slovenia Czech Republic Hong Kong Colombia Korea Australia
- Lot
- Lots 6JD079 and 6MD042
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States