ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c

Published April 12, 2017 · Recall initiated: February 23, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Worldwide Distribution - United States Nationwide: Denmark Sweden Norway Great Britain Belgium Netherlands Spain Israel Switzerland France Portugal Greece Germany Slovenia Czech Republic Hong Kong Colombia Korea Australia
Lot
Lots 6JD079 and 6MD042
Risk classification
Class III
Area (FDA)
device
Country
United States