Back to search
FDAFDA · Food and drugs·Recall number Z-1714-2026

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B

Published April 15, 2026 · Recall initiated: February 27, 2026

Summary

FDA records medical device Class I recall Z-1714-2026, published April 15, 2026 and initiated February 27, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 3120 kits; Lot: Medline SKU DYNJ42367B UDI/DI each 10193489211511, UDI/DI case 40193489211512, Lot Numbers: 24FBJ579, 24DBL828, 24DBE055, 24CBG967, 24ABN254, 23LBI540, 23KBF222, 23JBM761, 23JBA270, 23HBY791, 23HBD212, 23GBJ794, 23FBK893, 23EBQ313, 23DBC465. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medline Industries, LP
Distributor
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Lot
Medline SKU DYNJ42367B UDI/DI each 10193489211511, UDI/DI case 40193489211512, Lot Numbers: 24FBJ579, 24DBL828, 24DBE055, 24CBG967, 24ABN254, 23LBI540, 23KBF222, 23JBM761, 23JBA270, 23HBY791, 23HBD212, 23GBJ794, 23FBK893, 23EBQ313, 23DBC465.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: CVOR ANGIOGRAPHY PACK-LF, Medline SKU DYNJ42367B recalled?

Yes. FDA published recall Z-1714-2026 on April 15, 2026. Reported hazard: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit