ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-018-08

Published May 8, 2024 · Recall initiated: March 20, 2024

Summary

FDA records medical device Class II recall Z-1717-2024, published May 8, 2024 and initiated March 20, 2024. Recalling firm: Zimmer, Inc.; Status: Ongoing; Lot: UDI: (01)00889024088580(17)300112(10)64568169 (01)00889024088580(17)300307(10)64697763 (01)00889024088580(17)300225(10)64697764 (01)00889024088580(17)330215(10)65755769 (01)00889024088580(17)330218(10)65782566 (01)00889024088580(17)330211(10)65782567 (01)00889024088580(17)331208(10)66233085 Lot Numbers: 64568169 64697763 64697764 65755769 65782566 65782567 66233085. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Lot
UDI: (01)00889024088580(17)300112(10)64568169 (01)00889024088580(17)300307(10)64697763 (01)00889024088580(17)300225(10)64697764 (01)00889024088580(17)330215(10)65755769 (01)00889024088580(17)330218(10)65782566 (01)00889024088580(17)330211(10)65782567 (01)00889024088580(17)331208(10)66233085 Lot Numbers: 64568169 64697763 64697764 65755769 65782566 65782567 66233085
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-018-08 recalled?

Yes. FDA published recall Z-1717-2024 on May 8, 2024. Reported hazard: Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit