Back to search
FDAFDA · Food and drugs·Recall number Z-1719-2026

Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU # DYNJ68037C; 2. VEIN, Medline SKU # DYNJ89845; 3. VEIN, Medline SKU # DYNJ89845A.

Published April 15, 2026 · Recall initiated: February 27, 2026

Summary

FDA records medical device Class I recall Z-1719-2026, published April 15, 2026 and initiated February 27, 2026. Recalling firm: Medline Industries, LP; Status: Ongoing; Reported quantity: 2980 kits; Lot: Medline SKU # DYNJ68037C, UDI/DI each 10198459239328, UDI/DI case 40198459239329, Lot Number: 25HBV397; Medline SKU # DYNJ89845, UDI/DI each 10198459292460, UDI/DI case 40198459292461, Lot Number: 25CBU222; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26BBC195; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26ABI611; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25LBK402; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25KBH371; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25IBK975; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25HBQ426. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medline Industries, LP
Distributor
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Lot
Medline SKU # DYNJ68037C, UDI/DI each 10198459239328, UDI/DI case 40198459239329, Lot Number: 25HBV397; Medline SKU # DYNJ89845, UDI/DI each 10198459292460, UDI/DI case 40198459292461, Lot Number: 25CBU222; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26BBC195; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 26ABI611; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25LBK402; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25KBH371; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25IBK975; Medline SKU # DYNJ89845A, UDI/DI each 10198459490965, UDI/DI case 40198459490966, Lot Number: 25HBQ426.
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: 1. NON VASCULAR PACK, Medline SKU # DYNJ68037C; 2. VEIN, Medline SKU # DYNJ89845; 3. VEIN, Medline SKU # DYNJ89845A. recalled?

Yes. FDA published recall Z-1719-2026 on April 15, 2026. Reported hazard: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit