REF 154238 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size F6 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.

Published July 17, 2013 · Recall initiated: June 7, 2013

Summary

FDA records medical device Class II recall Z-1723-2013, published July 17, 2013 and initiated June 7, 2013. Recalling firm: Biomet, Inc.; Status: Terminated; Reported quantity: 25; Lot: M480960, M657120, M886450, M166890, M920650. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomet, Inc.
Distributor
Worldwide distribution: USA including NC, IN, PA, and OH; and countries of: Canada and Netherlands.
Lot
M480960, M657120, M886450, M166890, M920650
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was REF 154238 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size F6 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement. recalled?

Yes. FDA published recall Z-1723-2013 on July 17, 2013. Reported hazard: Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.

What should I do if I have this product?

Voluntary: Firm initiated · Letter