ZPLP Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-017-10
Published May 8, 2024 · Recall initiated: March 20, 2024
Summary
FDA records medical device Class II recall Z-1725-2024, published May 8, 2024 and initiated March 20, 2024. Recalling firm: Zimmer, Inc.; Status: Ongoing; Lot: GTIN: 00889024055896 Lot Numbers: Lot Number 64528516 64528517 64528518 64528522 64528523 64528527 64528528 64528529 64528530 64629028 64629030 64629031 64677270 64677272 64677276 64677278 64677279 64677281 64817167 64817168 64817169 64817170 64817171 64817172 64817173 64817174 64817175 64817176 65021088 65021089 65021090 65021091 65021092 65021093 65092268 65092269 65092270 65092271 65092272 65092273 65092274 65092275 65092276 65092277 65092278 65092279 65192604 65192605 65192606 65192607 65192608 65192609 66174613 66187674 66194640 66194643 66194646 66194649 66194652 66209353 66209363 66240704. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
- Lot
- GTIN: 00889024055896 Lot Numbers: Lot Number 64528516 64528517 64528518 64528522 64528523 64528527 64528528 64528529 64528530 64629028 64629030 64629031 64677270 64677272 64677276 64677278 64677279 64677281 64817167 64817168 64817169 64817170 64817171 64817172 64817173 64817174 64817175 64817176 65021088 65021089 65021090 65021091 65021092 65021093 65092268 65092269 65092270 65092271 65092272 65092273 65092274 65092275 65092276 65092277 65092278 65092279 65192604 65192605 65192606 65192607 65192608 65192609 66174613 66187674 66194640 66194643 66194646 66194649 66194652 66209353 66209363 66240704
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDAZPLP Distal Lateral Fibular Plate, Right, 8 Holes, 132 MM, Nonsterile- Indicated for temporary internalMay 8, 2024
- FDAZPLP Distal Lateral Fibular Plate, Left, 4 Holes, 80 MM, Nonsterile- Indicated for temporary internal fixationMay 8, 2024
Questions about this recall
Was ZPLP Distal Lateral Fibular Plate, Right, 10 Holes, 158 MM, Nonsterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-017-10 recalled?
Yes. FDA published recall Z-1725-2024 on May 8, 2024. Reported hazard: Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit