FDAFDA · Food and drugs·Recall number Z-1732-2013

CDI Blood Parameter Monitoring System 500 with Arterial and Venous blood parameter modules. Intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PCO2, PO2, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.

Published July 24, 2013 · Recall initiated: February 27, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Terumo discovered through its own testing that CDI 500 Blood Parameter Monitoring System software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV and the countries of Jordan, CANADA, AUSTRALIA, Japan ,Hong Kong ,UNITED ARAB EMIRATES (UAE), BELGIUM, South Korea, Malaysia, Mexico, Philippines, Singapore, Taiwan, and Thailand.
Lot
Catalog number: 500AV* and serial numbers: 0024, 0033, 0040, 0051, 1192, 1193, 1355,1356, 1420, 1423, 1464, 1465, 1467, 1468,1483,1504,1521-1524,1526,1527,1529,1530, 1602-1604, 1606, 1662, 2023, 2035, 2102-2107, 2800, 2901,2902, 3131-3135, 3261, 3262, 4132, 5330-5333, 5338-5340, 5578, and 5579.
Risk classification
Class II
Area (FDA)
device
Country
United States